| Type: | Package |
| Title: | Open Perimetry Interface |
| Version: | 3.0.5 |
| Date: | 2025-10-14 |
| Description: | Implementation of the Open Perimetry Interface (OPI) for simulating and controlling visual field machines using R. The OPI is a standard for interfacing with visual field testing machines (perimeters) first started as an open source project with support of Haag-Streit in 2010. It specifies basic functions that allow many visual field tests to be constructed. As of February 2022 it is fully implemented on the Haag-Streit Octopus 900 and 'CrewT ImoVifa' ('Topcon Tempo') with partial implementations on the Centervue Compass, Kowa AP 7000 and Android phones. It also has a cousin: the R package 'visualFields', which has tools for analysing and manipulating visual field data. |
| License: | Apache License (≥ 2) |
| URL: | https://opi.lei.org.au/ |
| Imports: | jsonlite, Rfast, abind, openssl |
| LazyData: | true |
| Encoding: | UTF-8 |
| RoxygenNote: | 7.3.1 |
| Collate: | opi.r Display.r PicoVR.r PhoneHMD.r ImoVifa.r dbTocd.r data-RtDbUnits.r data-RtSigmaUnits.r SimYes.r SimNo.r SimHenson.r SimGaussian.r zest.r mocs.r fourTwo.r full_threshold.r pix2deg.r QUESTP.r KTPsi.r Compass.r Octopus900.r Octopus600.r OPI-package.r SimHensonRT.r KowaAP7000Client.r MAIA.r |
| Suggests: | testthat (≥ 3.0.0) |
| Config/testthat/edition: | 3 |
| NeedsCompilation: | no |
| Packaged: | 2025-10-14 03:25:25 UTC; aturpin |
| Author: | Andrew Turpin [cre, aut, cph] (ORCID: 0000-0003-2559-8769), David Lawson [ctb, cph], Ivan Marin-Franch [ctb, cph], Matthias Muller [ctb], Jonathan Denniss [ctb, cph], Astrid Zeman [ctb], Giovanni Montesano [ctb] |
| Maintainer: | Andrew Turpin <andrew.turpin@lei.org.au> |
| Repository: | CRAN |
| Date/Publication: | 2025-10-14 04:50:02 UTC |
OPI: Open Perimetry Interface
Description
Implementation of the Open Perimetry Interface (OPI) for simulating and controlling visual field machines using R. The OPI is a standard for interfacing with visual field testing machines (perimeters) first started as an open source project with support of Haag-Streit in 2010. It specifies basic functions that allow many visual field tests to be constructed. As of February 2022 it is fully implemented on the Haag-Streit Octopus 900 and 'CrewT ImoVifa' ('Topcon Tempo') with partial implementations on the Centervue Compass, Kowa AP 7000 and Android phones. It also has a cousin: the R package 'visualFields', which has tools for analysing and manipulating visual field data.
Author(s)
Maintainer: Andrew Turpin andrew.turpin@lei.org.au (ORCID: 0000-0003-2559-8769) [copyright holder]
Other contributors:
David Lawson [contributor, copyright holder]
Ivan Marin-Franch [contributor, copyright holder]
Matthias Muller [contributor]
Jonathan Denniss [contributor, copyright holder]
Astrid Zeman [contributor]
Giovanni Montesano [contributor]
See Also
Useful links:
Global environment for OPI to hold machine specific constants, etc.
Description
Global environment for OPI to hold machine specific constants, etc.
Usage
.opi_env
Format
An object of class environment of length 11.
Full Threshold
Description
FT begins with a 4-2dB staircase beginning at level
est. If the final estimate (last seen) is more than 4dB away
from est, a second 4-2 staircase is completed beginning at the
estimate returned from the first
Usage
FT(est = 25, instRange = c(0, 40), verbose = FALSE, makeStim, ...)
FT.start(est = 25, instRange = c(0, 40), makeStim, ...)
FT.step(state, nextStim = NULL)
FT.stop(state)
FT.final(state)
Arguments
est |
Starting estimate in dB |
instRange |
Dynamic range of the instrument c(min,max) in dB |
verbose |
True if you want each presentation printed |
makeStim |
A function that takes a dB value and numPresentations and returns an OPI datatype ready for passing to opiPresent |
... |
Extra parameters to pass to the opiPresent function |
state |
Current state of the FT returned by |
nextStim |
A valid object for |
Details
This is an implementation of a 4-2 1-up 1-down staircase as
implemented in the first Humphrey Field Analyzer. The initial staircase
starts at est and proceeds in steps of 4 dB until the first
reversal, and 2dB until the next reversal. The last seen stimulus is
taken as the threshold value. If, after the first staircase, the
threshold is more than 4 dB away from the starting point, then a second
staircase is initiated with a starting point equal to the threshold
found with the first staircase.
Note this function will repeatedly call opiPresent for a stimulus
until opiPresent returns NULL (ie no error occurred)
If more than one FT is to be interleaved (for example, testing multiple
locations), then the FT.start, FT.step, FT.stop
and FT.final calls can maintain the state of the FT after each
presentation, and should be used. If only a single FT is required, then
the simpler FT can be used. See examples below
Value
Single location
Returns a list containing
-
npresTotal number of presentations -
respSeqResponse sequence stored as a list of (seen,dB) pairs -
firstFirst staircase estimate in dB -
finalFinal threshold estimate in dB
Multiple locations
FT.start returns a list that can be passed to FT.step,
FT.stop, and FT.final. It represents the state of a FT
at a single location at a point in time and contains the following.
-
nameFT A copy of all of the parameters supplied to FT.start:
startingEstimate=est,minStimulus=instRange[1],maxStimulus=instRange[2],makeStim, andopiParams=list(...).-
currentLevelThe next stimulus to present. -
lastSeenThe last seen stimulus. -
lastResponseThe last response given. -
firstStairResultThe result of the first staircase (initiallyNA). -
secondStairResultThe result of the first staircase (initiallyNA, and could remainNA). -
finishedTRUEif staircase has finished (2 reversals, or max/min seen/not-seen twice). -
numberOfReversalsNumber of reversals so far. -
currSeenLimitNumber of timesmaxStimulushas been seen. -
currNotSeenLimitNumber of timesminStimulusnot seen. -
numPresentationsNumber of presentations so far. -
stimuliVector of stimuli shown at each call toFT.step. -
responsesVector of responses received (1 seen, 0 not) received at each call toFT.step. -
responseTimesVector of response times received at each call toFT.step.
FT.step returns a list containing
-
stateThe new state after presenting a stimuli and getting a response. -
respThe return from theopiPresentcall that was made.
FT.stop returns TRUE if the first staircase has had 2 reversals, or
maxStimulus is seen twice or minStimulus is not seen twice and the
final estimate is within 4 dB of the starting stimulus. Returns TRUE if
the second staircase has had 2 reversals, or maxStimulus is seen twice or
minStimulus is not seen twice
FT.final returns the final estimate of threshold based on state, which is
the last seen in the second staircase, if it ran, or the first staircase otherwise
FT.final.details returns a list containing
-
finalThe final threshold. -
firstThe threshold determined by the first staircase (might be different from final). -
stopReasonEitherReversals,Max, orMinwhich are the three ways in which FT can terminate. -
npNumber of presentation for the whole procedure (including both staircases if run).
References
A. Turpin, P.H. Artes and A.M. McKendrick. "The Open Perimetry Interface: An enabling tool for clinical visual psychophysics", Journal of Vision 12(11) 2012.
H. Bebie, F. Fankhauser and J. Spahr. "Static perimetry: strategies", Acta Ophthalmology 54 1976.
C.A. Johnson, B.C. Chauhan, and L.R. Shapiro. "Properties of staircase procedures for estimating thresholds in automated perimetry", Investigative Ophthalmology and Vision Science 33 1993.
See Also
dbTocd, opiPresent, fourTwo.start
Examples
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500)
class(s) <- "opiStaticStimulus"
return(s)
}
chooseOpi("SimHenson")
if (!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
result <- FT(makeStim=makeStim, tt=30, fpr=0.15, fnr=0.01)
if (!is.null(opiClose()$err))
warning("opiClose() failed")
##############################################
# This section is for multiple FTs
##############################################
chooseOpi("SimHenson")
if (!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
makeStimHelper <- function(db,n, x, y) { # returns a function of (db,n)
ff <- function(db, n) db+n
body(ff) <- substitute({
s <- list(x=x, y=y, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500)
class(s) <- "opiStaticStimulus"
return(s)
}, list(x=x,y=y))
return(ff)
}
# List of (x, y, true threshold) triples
locations <- list(c(9,9,30), c(-9,-9,32), c(9,-9,31), c(-9,9,33))
# Setup starting states for each location
states <- lapply(locations, function(loc) {
FT.start(makeStim=makeStimHelper(db,n,loc[1],loc[2]),
tt=loc[3], fpr=0.03, fnr=0.01)})
# Loop through until all states are "stop"
while(!all(st <- unlist(lapply(states, FT.stop)))) {
i <- which(!st) # choose a random,
i <- i[runif(1, min=1, max=length(i))] # unstopped state
r <- FT.step(states[[i]]) # step it
states[[i]] <- r$state # update the states
}
finals <- lapply(states, FT.final) # get final estimates of threshold
for(i in 1:length(locations)) {
cat(sprintf("Location (%+2d,%+2d) ",locations[[i]][1], locations[[i]][2]))
cat(sprintf("has threshold %4.2f\n", finals[[i]]))
}
if(!is.null(opiClose()$err))
warning("opiClose() failed")
An implementation of Kontsevich and Tyler \Psi algorithm.
Description
An implementation of Kontsevich and Tyler (Vis Res 39 (1999) pages 2729–2737 default parameterised for Standard Automated Perimetry. based on A. Turpin, D. Jankovic and A.M. McKendrick, "Identifying Steep Psychometric Function Slope Quickly in Clinical Applications", Vision Research, 50(23). November 2010. Pages 2476-2485
Usage
KTPsi(
domains = list(slopes = 1:5, thresholds = 20:40, fps = c(0, 0.025, 0.05, 0.1, 0.2), fns
= c(0, 0.025, 0.05, 0.1, 0.2)),
priors = list(slopes = rep(1, length(domains$slopes))/length(domains$slopes),
thresholds = rep(1, length(domains$thresholds))/length(domains$thresholds), fps =
rep(1, length(domains$fps))/length(domains$fps), fns = rep(1,
length(domains$fns))/length(domains$fns)),
stimValues = 17:40,
stopType = "N",
stopValue = 140,
maxPresentations = 200,
minInterStimInterval = NA,
maxInterStimInterval = NA,
verbose = 0,
makeStim,
...
)
KTPsi.start(
domains = list(slopes = 1:5, thresholds = 20:40, fps = c(0, 0.025, 0.05, 0.1, 0.2), fns
= c(0, 0.025, 0.05, 0.1, 0.2)),
priors = list(slopes = rep(1, length(domains$slopes))/length(domains$slopes),
thresholds = rep(1, length(domains$thresholds))/length(domains$thresholds), fps =
rep(1, length(domains$fps))/length(domains$fps), fns = rep(1,
length(domains$fns))/length(domains$fns)),
stimValues = 17:40,
stopType = "N",
stopValue = 140,
maxPresentations = 200,
minInterStimInterval = NA,
maxInterStimInterval = NA,
verbose = 0,
makeStim,
...
)
KTPsi.step(state, nextStim = NULL, fixedStimValue = NA)
KTPsi.final(state, method = "expectation")
KTPsi.stop(state)
Arguments
domains |
A list of 4 vectors:
|
priors |
A list of 4 vectors:
Each prior should the same length as its |
stimValues |
Vector of allowable stimulus values. |
stopType |
|
stopValue |
Value for number of presentations ( |
maxPresentations |
Maximum number of presentations regardless of |
minInterStimInterval |
If both |
maxInterStimInterval |
|
verbose |
|
makeStim |
A function that takes a stimulus value and numPresentations and returns an OPI datatype ready for passing to opiPresent. See examples. |
... |
Extra parameters to pass to the opiPresent function |
state |
Current state of the KTPsi as returned by (eg) |
nextStim |
The next stimulus to present in a suitable format for passing to |
fixedStimValue |
Currently ignored. |
method |
Either |
Details
The assumed psychometric function is the cumulative Gaussian:
\mbox{fp}+(1-\mbox{fp}-\mbox{fn})(1-\mbox{pnorm}(x, \mbox{threshold}, \mbox{slope})
hence domain$slopes are standard deviations and domain$thresholds are the mean.
While it is assumed that domains$thresholds and stimValues are in dB, this need not be the case.
As long as the makeStim function converts stimValues into cd/\mbox{m}^2
for the opiPresent function, then any units should work.
The checkFixationOK function is called (if present in stim made from makeStim)
after each presentation, and if it returns FALSE, the pdf for that state is not changed
(ie the presentation is ignored), but the stim, number of presentations etc is recorded in
the state.
If more than one KTPsi is to be interleaved (for example, testing multiple locations), then thePsi
KTPsi.start, KTPsi.step, KTPsi.stop and KTPsi.final calls can maintain
the state of the KTPsi after each presentation, and should be used. If only a single KTPsi is
required, then the simpler KTPsi function can be used, which is a wrapper for the four functions
that maintain state. See examples below.
Value
Single location
KTPsi returns a list containing
-
npresTotal number of presentations used. -
respSeqResponse sequence stored as a matrix: row 1 is dB values of stimuli, row 2 is 1/0 for seen/not-seen, row 3 is fixated 1/0 (always 1 ifcheckFixationOKnot present in stim objects returned frommakeStim). -
pdfsIfverboseis bigger than 0, then this is a list of the pdfs used for each presentation, otherwise NULL. -
finalThe mean/median/mode of the final pdf, depending onstimChoice, which is the determined threshold. -
opiRespA list of responses received from each successful call toopiPresentwithinKTPsi.
Multilple locations
KTPsi.start returns a list that can be passed to KTPsi.step, KTPsi.stop, and KTPsi.final. It represents the state of a KTPsi at a single location at a point in time and contains the following.
-
nameKTPsi A copy of all of the parameters supplied to KTPsi.start:
domains,priors,stimValues,stopType,stopValue,maxPresentations,makeStimandopiParams.-
psiA matrix wherepsi[domain_index, stim]is the probability of seeingstimassuming the psychometric function for the domain indexdomain_index. -
labelsA text representation ofpsi[domain_index, ], or the the psychometric function for the domain indexdomain_index. -
pdfCurrent pdf: vector of probabilities the same length as product of lengths ofdomainelements. -
numPresentationsThe number of timesKTPsi.stephas been called on this state. -
stimuliA vector containing the stimuli used at each call ofKTPsi.step. -
responsesA vector containing the responses received at each call ofKTPsi.step. -
responseTimesA vector containing the response times received at each call ofKTPsi.step. -
fixatedA vector containing TRUE/FALSE if fixation was OK according tocheckFixationOKfor each call ofKTPsi.step(defaults to TRUE ifcheckFixationOKnot present). -
opiRespA list of responses received from each call toopiPresentwithinKTPsi.step.
KTPsi.step returns a list containing
-
stat:The new state after presenting a stimuli and getting a response. -
respThe return from theopiPresentcall that was made.
KTPsi.stop returns TRUE if the KTPsi has reached its stopping criteria, and FALSE otherwise.
KTPsi.final returns an estimate of threshold based on state based on its parameter.
TRUE if the state has reached its stopping criteria, and FALSE otherwise.
References
Kontsevich and Tyler. Vision Research 39 (1999) pages 2729–2737.
A. Turpin, D. Jankovic and A.M. McKendrick, "Identifying Steep Psychometric Function Slope Quickly in Clinical Applications", Vision Research, 50(23). November 2010. Pages 2476-2485
A. Turpin, P.H. Artes and A.M. McKendrick "The Open Perimetry Interface: An enabling tool for clinical visual psychophysics", Journal of Vision 12(11) 2012.
See Also
Examples
chooseOpi("SimGaussian")
if(!is.null(opiInitialize(sd = 2)$err))
stop("opiInitialize failed")
# This section is for single location KTPsi
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}
KTPsi(makeStim = makeStim, stopType="H", stopValue= 3, tt=30, fpr=0.03)
KTPsi(makeStim = makeStim, stopType="N", stopValue= 27, verbose = 0, tt=30, fpr=0.03)
# For multiple locations...
## Not run:
states <- lapply(1:10, function(loc) KTPsi.start(makeStim = makeStim))
unfinished <- 1:10
while (length(unfinished) > 0) {
loc <- unfinished[[1]]
states[[loc]] <- KTPsi.step(states[[loc]])$state
if (KTPsi.stop(states[[loc]]))
unfinished <- unfinished[-1]
}
## End(Not run)
Method of Constant Stimuli (MOCS)
Description
MOCS performs either a yes/no or n-interval-forced-choice Method of Constant Stimuli test
Usage
MOCS(
params = NA,
order = "random",
responseWindowMeth = "constant",
responseFloor = 1500,
responseHistory = 5,
keyHandler = function(correct, ret) return(list(seen = TRUE, time = 0, err = NULL)),
interStimMin = 200,
interStimMax = 500,
beep_function,
makeStim,
stim_print,
...
)
Arguments
params |
A matrix where each row is
|
order |
Control the order in which the stimuli are presented.
|
responseWindowMeth |
Control time perimeter waits for response.
|
responseFloor |
Minimum response window (for any |
responseHistory |
Number of past yeses to average to get response window
(only used if |
keyHandler |
Function to get a keyboard input and returns as for |
interStimMin |
Regardless of response, wait |
interStimMax |
Regardless of response, wait |
beep_function |
A function that takes the string |
makeStim |
A helper function to take a row of |
stim_print |
A function that takes an |
... |
Extra parameters to pass to the opiPresent function. |
Details
Whether the test is yes/no or forced-choice is determined by the number of columns
in params. The code simply presents all columns from 5 onwards and collects a
response at the end. So if there is only 5 columns, it is a yes/no task. If there are 6
columns it is a 2-interval-forced-choice. Generally, an nIFC experiment has 4+n columns in
params.
Note that when the order is "random", the number of trials in the test will be
the sum of the 3rd column of params. When the order is "fixed", there is
only one presentation per row, regardless of the value in the 3rd column of params.
If a response is received before the final trial in a nIFC experiment, it is ignored.
If the checkFixationOK function is present in a stimulus, then it is called after each
presentation, and the result is “anded” with each stimulus in a trial to get a TRUE/FALSE
for fixating on all stimuli in a trial.
Value
Returns a data.frame with one row per stimulus copied from params with extra columns
appended: checkFixation checks, and the return values from opiPresent()
(see example). These last values will differ depending on which
machine/simulation you are running (as chosen with chooseOpi().
column 1: x
column 2: y
column 3: location number
column 4: number of times to repeat this stim
column 5: correct stimulus index
column 6: TRUE/FALSE was fixating for all presentations in this trial according to
checkFixationOKcolumn 7...: columns from params
...: columns from opiPresent return
References
A. Turpin, P.H. Artes and A.M. McKendrick. "The Open Perimetry Interface: An enabling tool for clinical visual psychophysics", Journal of Vision 12(11) 2012.
See Also
Examples
# For the Octopus 900
# Check if pupil centre is within 10 pixels of (160,140)
checkFixationOK <- function(ret) return(sqrt((ret$pupilX - 160)^2 + (ret$pupilY - 140)^2) < 10)
# Return a list of opi stim objects (list of class opiStaticStimulus) for each level (dB) in
# p[5:length(p)]. Each stim has responseWindow BETWEEN_FLASH_TIME, except the last which has
# rwin. This one assumes p is on old Octopus 900 dB scale (0dB == 4000 cd/m^2).
makeStim <- function(p, rwin) {
BETWEEN_FLASH_TIME <- 750 # ms
res <- NULL
for(i in 5:length(p)) {
s <- list(x=p[1], y=p[2], level=dbTocd(p[i],4000/pi), size=0.43, duration=200,
responseWindow=ifelse(i < length(p), BETWEEN_FLASH_TIME, rwin),
checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
res <- c(res, list(s))
}
return(res)
}
################################################################
# Read in a key press 'z' is correct==1, 'm' otherwise
# correct is either 1 or 2, whichever is the correct interval
#
# Return list(seen={TRUE|FALSE}, time=time, err=NULL))
# seen is TRUE if correct key pressed
################################################################
## Not run:
if (length(dir(".", "getKeyPress.py")) < 1)
stop('Python script getKeyPress.py missing?')
## End(Not run)
keyHandler <- function(correct, ret) {
return(list(seen=TRUE, time=0, err=NULL))
ONE <- "b'z'"
TWO <- "b'm'"
time <- Sys.time()
key <- 'q'
while (key != ONE && key != TWO) {
a <- system('python getKeyPress.py', intern=TRUE)
key <- a # substr(a, nchar(a), nchar(a))
print(paste('Key pressed: ',key,'from',a))
if (key == "b'8'")
stop('Key 8 pressed')
}
time <- Sys.time() - time
if ((key == ONE && correct == 1) || (key == TWO && correct == 2))
return(list(seen=TRUE, time=time, err=NULL))
else
return(list(seen=FALSE, time=time, err=NULL))
}
################################################################
# Read in return value from opipresent with F310 controller.
# First param is correct, next is 1 for left button, 2 for right button
# Left button (LB) is correct for interval 1, RB for interval 2
# correct is either 1 or 2, whichever is the correct interval
#
# Return list(seen={TRUE|FALSE}, time=time, err=NULL))
# seen is TRUE if correct key pressed
################################################################
F310Handler <- function(correct, opiResult) {
z <- opiResult$seen == correct
opiResult$seen <- z
return(opiResult)
}
################################################################
# 2 example beep_function
################################################################
## Not run:
require(beepr)
myBeep <- function(type='None') {
if (type == 'correct') {
beepr::beep(2) # coin noise
Sys.sleep(0.5)
}
if (type == 'incorrect') {
beepr::beep(1) # system("rundll32 user32.dll,MessageBeep -1") # system beep
#Sys.sleep(0.0)
}
}
require(audio)
myBeep <- function(type="None") {
if (type == 'correct') {
wait(audio::play(sin(1:10000/10)))
}
if (type == 'incorrect') {
wait(audio::play(sin(1:10000/20)))
}
}
## End(Not run)
################################################################
# An example stim_print function
################################################################
## Not run:
stim_print <- function(s, ret) {
sprintf("%4.1f %2.0f",cdTodb(s$level,10000/pi), ret$seen)
}
## End(Not run)
QUEST+
Description
An implementation of the Bayesian test procedure QUEST+ by AB Watson. This is mostly a translation of the MATLAB implementation by P Jones (see References). Its use is similar to ZEST. The objective is to estimate parameters of a function that defines the probability of responding stimuli. The steps are optimized based on entropy rather than the mean or the mode of the current pdfs.
Usage
QUESTP(
Fun,
stimDomain,
paramDomain,
likelihoods = NULL,
priors = NULL,
stopType = "H",
stopValue = 4,
maxSeenLimit = 2,
minNotSeenLimit = 2,
minPresentations = 1,
maxPresentations = 100,
minInterStimInterval = NA,
maxInterStimInterval = NA,
verbose = 0,
makeStim,
...
)
QUESTP.Prior(state, priors = NULL)
QUESTP.Likelihood(state)
QUESTP.start(
Fun,
stimDomain,
paramDomain,
likelihoods = NULL,
priors = NULL,
stopType = "H",
stopValue = 4,
maxSeenLimit = 2,
minNotSeenLimit = 2,
minPresentations = 1,
maxPresentations = 100,
makeStim,
...
)
getTargetStim(state)
QUESTP.step(state, nextStim = NULL)
QUESTP.stop(state)
QUESTP.final(state, Choice = "mean")
QUESTP.stdev(state, WhichP = NULL)
QUESTP.entropy(state)
Arguments
Fun |
Function to be evaluated, of the form |
stimDomain |
Domain of values for the stimulus. Can be multi-dimensional (list, one element per dimension) |
paramDomain |
Domain of values for pdfs of the parameters in Fun. Can be multi-parametric (list, one element per parameter). |
likelihoods |
Pre-computed likelihoods if available (for QUESTP.start) |
priors |
Starting probability distributions for the parameter domains (list, one element per parameter) |
stopType |
|
stopValue |
Value for number of presentations ( |
maxSeenLimit |
Will terminate if |
minNotSeenLimit |
Will terminate if |
minPresentations |
Minimum number of presentations |
maxPresentations |
Maximum number of presentations regarless of |
minInterStimInterval |
If both |
maxInterStimInterval |
|
verbose |
|
makeStim |
A function that takes a dB value and numPresentations and returns an OPI datatype ready for passing to opiPresent. See examples. |
... |
Extra parameters to pass to the opiPresent function |
state |
Current state of the QUESTP returned by |
nextStim |
A valid object for |
Choice |
How to compute final values in QUESTP.final ("mean","mode","median") |
WhichP |
Which parameter (numeric index) to monitor when calling QUESTP.stdev directly (returns max(stdev) if unspecified) |
Details
An implementation of the Bayesian test procedure QUEST+ by AB Watson. This is mostly a translation of the MATLAB implementation by P Jones (see References). Its use is similar to ZEST. The objective is to estimate parameters of a function that defines the probability of responding to stimuli. The steps are optimized based on entropy rather than the mean or the mode of the current pdfs.
The stimulus, parameter and response domain are separate and can be multidimensional. Each parameter has its own pdf. For evaluation, the pdfs are chained as a long vector (so no co-variances are considered). More complex functions will require larger combined pdfs and longer computation time for likelihoods and updating at each step. In these cases, it is recommended to pre-calculate the likelihoods using QUESTP.Likelihood and store them.
The function to be fitted needs to output a probability of seen (i.e. pseen = function(stim, param){...})
and must take stim and param as inputs. stim is a vector with length = number of stimulus dimensions
(in simple one-dimensional cases, the intensity in dB). param is a vector with length = number
of parameters to be fitted in Fun.
For example, QUEST+ can fit a Gaussian psychometric function with stimDomain = {0, 1,..., 39, 40} dB
and paramDomain = ({0, 1,..., 39, 40}; {0.5, 1,..., 5.5, 6}) dB for the mean and
standard deviation respectively. A standard ZEST procedure can be replicated by setting
stimDomain = {0, 1,..., 39, 40} dB and paramDomain = ({0, 1,..., 39, 40}; {1}) dB, i.e. by setting the
stimDomain = paramDomain for the mean and by having a static standard deviation = 1 dB. Note however that
the stimulus selection is always based on entropy and not on the mean/mode of the current pdf. See examples below
Note this function will repeatedly call opiPresent for a stimulus until opiPresent
returns NULL (ie no error occurred).
The checkFixationOK function is called (if present in stim made from makeStim)
after each presentation, and if it returns FALSE, the pdf for that location is not changed
(ie the presentation is ignored), but the stim, number of presentations etc is recorded in
the state.
If more than one QUESTP is to be interleaved (for example, testing multiple locations), then the
QUESTP.start, QUESTP.step, QUESTP.stop and QUESTP.final calls can maintain
the state of the QUESTP after each presentation, and should be used. If only a single QUESTP is
required, then the simpler QUESTP can be used, which is a wrapper for the four functions
that maintain state. See examples below.
Value
##Single location
QUESTP returns a list containing
-
npresTotal number of presentations used. -
respSeqResponse sequence stored as a data frame: column 1 is a string identified of a (potentially) multidimensional stimulus values of stimuli (dimensions chained into a string), column 2 is 1/0 for seen/not-seen, column 3 is fixated 1/0 (always 1 ifcheckFixationOKnot present in stim objects returned frommakeStim). All additional columns report each stimulus dimension, one for each row. -
pdfsIfverboseis bigger than 0, then this is a list of the pdfs used for each presentation, otherwise NULL. -
finalThe mean (default, strongly suggested)/median/mode of the parameters' pdf, depending onChoice. -
opiRespA list of responses received from each successful call toopiPresentwithinQUESTP.
Multiple locations
QUESTP.start returns a list that can be passed to QUESTP.step, QUESTP.stop, and
QUESTP.final. It represents the state of a QUESTP at a single location at a point in time
and contains the following.
-
nameQUESTP A copy of all of the parameters supplied to QUESTP.start:
stimDomain,paramDomain,likelihoods,priors,stopType,stopValue,maxSeenLimit,minNotSeenLimit,minPresentations,maxPresentations,makeStim, andopiParams.-
pdfCurrent pdf: vector of probabilities, collating all parameter domains. -
priorsPList of starting pdfs, one for each parameter. -
numPresentationsThe number of timesQUESTP.stephas been called on this state. -
stimuliA vector containing the stimuli used at each call ofQUESTP.step. -
responsesA vector containing the responses received at each call ofQUESTP.step. -
responseTimesA vector containing the response times received at each call ofQUESTP.step. -
fixatedA vector containing TRUE/FALSE if fixation was OK according tocheckFixationOKfor each call ofQUESTP.step(defaults to TRUE ifcheckFixationOKnot present). -
opiRespA list of responses received from each call toopiPresentwithinQUESTP.step.
QUESTP.step returns a list containing
-
stateThe new state after presenting a stimuli and getting a response. -
respThe return from theopiPresentcall that was made.
QUESTP.stop returns TRUE if the QUESTP has reached its stopping criteria, and FALSE otherwise.
QUESTP.final returns an estimate of parameters based on state. If state$Choice
is mean then the mean is returned (the only one that really makes sense for QUESTP).
If state$Choice is mode then the
mode is returned. If state$Choice is median then the median is returned.
References
Andrew B. Watson; QUEST+: A general multidimensional Bayesian adaptive psychometric method. Journal of Vision 2017;17(3):10. doi: https://doi.org/10.1167/17.3.10.
Jones, P. R. (2018). QuestPlus: a MATLAB implementation of the QUEST+ adaptive psychometric method, Journal of Open Research Software, 6(1):27. doi: http://doi.org/10.5334/jors.195
A. Turpin, P.H. Artes and A.M. McKendrick "The Open Perimetry Interface: An enabling tool for clinical visual psychophysics", Journal of Vision 12(11) 2012.
See Also
Examples
chooseOpi("SimHenson")
if(!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
#########################################################
# This section is for single location QUESTP
# This example fits a FoS curve
# Note: only fitting threshold and slope,
# modify the domain for FPR and FNR to fit those as well
#########################################################
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}
#True parameters (variability is determined according to Henson et al. based on threshold)
loc <- list(threshold = 20, fpr = 0.05, fnr = 0.05)
#Function to fit (Gaussian psychometric function)
pSeen <- function(x, params){return(params[3] +
(1 - params[3] - params[4]) *
(1 - pnorm(x, params[1], params[2])))}
#QUEST+
QP <- QUESTP(Fun = pSeen,
stimDomain = list(0:50),
paramDomain = list(seq(0, 40, 1), #Domain for the 50% threshold (Mean)
seq(.5, 8, .5), #Domain for the slope (SD)
seq(0.05, 0.05, 0.05), #Domain for the FPR (static)
seq(0.05, 0.05, 0.05)), #Domain for the FNR (static)
stopType="H", stopValue=4, maxPresentations=500,
makeStim = makeStim,
tt=loc$threshold, fpr=loc$fpr, fnr=loc$fnr,
verbose = 2)
#Plots results
#Henson's FoS function (as implemented in OPI - ground truth)
HensFunction <- function(Th){
SD <- exp(-0.081*Th + 3.27)
SD[SD > 6] <- 6
return(SD)
}
#Stimulus domain
dB_Domain <- 0:50
FoS <- pSeen(dB_Domain, params = QP$final) # Estimated FoS
FoS_GT <- pSeen(dB_Domain, params = c(loc$threshold, HensFunction(loc$threshold),
loc$fpr, loc$fnr)) #Ground truth FoS (based on Henson et al.)
#Plot (seen stimuli at the top, unseen stimuli at the bottom)
plot(dB_Domain, FoS_GT, type = "l", ylim = c(0, 1), xlab = "dB", ylab = "% seen", col = "blue")
lines(dB_Domain, FoS, col = "red")
points(QP$respSeq$stimuli, QP$respSeq$responses, pch = 16,
col = rgb(red = 1, green = 0, blue = 0, alpha = 0.1))
legend("top", inset = c(0, -.2),legend = c("True","Estimated","Stimuli"),
col=c("blue", "red","red"), lty=c(1,1,0),
pch = c(16, 16, 16), pt.cex = c(0, 0, 1),
horiz = TRUE, xpd = TRUE, xjust = 0)
if (!is.null(opiClose()$err))
warning("opiClose() failed")
chooseOpi("SimHenson")
if(!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
######################################################################
# This section is for single location QUESTP
# This example shows that QUEST+ can replicate a ZEST procedure
# by fitting a FoS curve with fixed Slope, FPR and FNR
# Compared with ZEST
# Note that QUEST+ should be marginally more efficient in selecting
# the most informative stimulus
######################################################################
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}
#True parameters (variability is determined according to Henson et al. based on threshold)
loc <- list(threshold = 30, fpr = 0.03, fnr = 0.03)
#Function to fit (Gaussian psychometric function - Fixed slope (same as default likelihood in ZEST))
pSeen <- function(domain, tt){{0.03+(1-0.03-0.03)*(1-pnorm(domain,tt,1))}}
# ZEST-like QUEST+ procedure
QP <- QUESTP(Fun = pSeen,
stimDomain = list(0:40),
paramDomain = list(seq(0, 40, 1)),
stopType="S", stopValue=1.5, maxPresentations=500,
makeStim = makeStim,
tt=loc$threshold, fpr=loc$fpr, fnr=loc$fnr,
verbose = 2)
# ZEST
ZE <- ZEST(domain = 0:40,
stopType="S", stopValue=1.5, maxPresentations=500,
makeStim = makeStim,
tt=loc$threshold, fpr=loc$fpr, fnr=loc$fnr,
verbose = 2)
#Plots results
#Henson's FoS function (as implemented in OPI - ground truth)
HensFunction <- function(Th){
SD <- exp(-0.081*Th + 3.27)
SD[SD > 6] <- 6
return(SD)
}
#Stimulus domain
dB_Domain <- 0:50
FoS_QP <- pSeen(domain = dB_Domain, tt = QP$final) # Estimated FoS
FoS_ZE <- pSeen(domain = dB_Domain, tt = ZE$final) # Estimated FoS
#Plot (seen stimuli at the top, unseen stimuli at the bottom)
plot(dB_Domain, FoS_QP, type = "l", ylim = c(0, 1), xlab = "dB", ylab = "% seen", col = "blue")
lines(dB_Domain, FoS_ZE, col = "red")
points(QP$respSeq$stimuli, QP$respSeq$responses, pch = 16,
col = rgb(red = 0, green = 0, blue = 1, alpha = 0.5))
points(ZE$respSeq[1,], ZE$respSeq[2,], pch = 16,
col = rgb(red = 1, green = 0, blue = 0, alpha = 0.5))
legend("bottomleft", legend = c("QUEST+","ZEST","Stimuli QUEST+", "Stimuli ZEST"),
col=c("blue", "red","blue","red"), lty=c(1,1,0,0),
pch = c(16, 16, 16, 16), pt.cex = c(0, 0, 1, 1),
horiz = FALSE, xpd = TRUE, xjust = 0)
abline(v = loc$threshold, lty = "dashed")
if (!is.null(opiClose()$err))
warning("opiClose() failed")
chooseOpi("SimHenson")
if(!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
#########################################################
# This section is for single location QUESTP
# This example fits a broken stick spatial summation function
# with a multi-dimensional stimulus (varying in size and intensity).
# Stimulus sizes are limited to GI, GII, GIII, GIV and GV.
# The example also shows how to use a helper function to
# simulate responses to multi-dimensional stimuli
# (here, the simulated threshold varies based on stimulus size)
#########################################################
makeStim <- function(stim, n) {
s <- list(x=9, y=9, level=dbTocd(stim[1]), size=stim[2], color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}
# Helper function for true threshold (depends on log10(stimulus size),
# diameter assumed to be the second element of stim vector)
ttHelper_SS <- function(location) { # returns a function of (stim)
ff <- function(stim) stim
body(ff) <- substitute(
{return(SensF(log10(pi*(stim[2]/2)^2), c(location$Int1, location$Int2, location$Slo2)))}
)
return(ff)
}
# Function of sensivity vs SSize (log10(stimulus area))
SensF <- function(SSize, params){
Sens <- numeric(length(SSize))
for (i in 1:length(SSize)){
Sens[i] <- min(c(params[1] + 10*SSize[i], params[2] + params[3]*SSize[i]))
}
Sens[Sens < 0] <- 0
return(Sens)
}
Sizes <- c(0.1, 0.21, 0.43, 0.86, 1.72)
#True parameters (variability is determined according to Henson et al. based on threshold)
loc <- list(Int1 = 32, Int2 = 28, Slo2 = 2.5, fpr = 0.05, fnr = 0.05, x = 9, y = 9)
# Function to fit (probability of seen given a certain stimulus intensity and size,
# for different parameters)
pSeen <- function(stim, params){
Th <- SensF(log10(pi*(stim[2]/2)^2), params)
return(0.03 +
(1 - 0.03 - 0.03) *
(1 - pnorm(stim[1], Th, 1)))
}
## Not run:
set.seed(111)
#QUEST+ - takes some time to calculate likelihoods
QP <- QUESTP(Fun = pSeen,
stimDomain = list(0:50, Sizes),
paramDomain = list(seq(0, 40, 1), # Domain for total summation intercept
seq(0, 40, 1), # Domain for partial summation intercept
seq(0, 3, 1)), # Domain for partial summation slope
stopType="H", stopValue=1, maxPresentations=500,
makeStim = makeStim,
ttHelper=ttHelper_SS(loc), tt = 30,
fpr=loc$fpr, fnr=loc$fnr,
verbose = 2)
#Stimulus sizes
G <- log10(c(pi*(0.1/2)^2, pi*(0.21/2)^2, pi*(0.43/2)^2, pi*(0.86/2)^2, pi*(1.72/2)^2));
SizesP <- seq(min(G), max(G), .05)
# True and estimated response
Estim_Summation <- SensF(SizesP, params = QP$final) # Estimated spatial summation
GT_Summation <- SensF(SizesP, params = c(loc$Int1, loc$Int2, loc$Slo2)) # True spatial summation
#Plot
plot(10^SizesP, GT_Summation, type = "l", ylim = c(0, 40), log = "x",
xlab = "Stimulus area (deg^2)", ylab = "Sensitivity (dB)", col = "blue")
lines(10^SizesP, Estim_Summation, col = "red")
points(pi*(QP$respSeq$stimuli.2/2)^2, QP$respSeq$stimuli.1, pch = 16,
col = rgb(red = 1, green = 0, blue = 0, alpha = 0.3))
legend("top", inset = c(0, -.2),legend = c("True","Estimated","Stimuli"),
col=c("blue", "red","red"), lty=c(1,1,0),
pch = c(16, 16, 16), pt.cex = c(0, 0, 1),
horiz = TRUE, xpd = TRUE, xjust = 0)
## End(Not run)
if (!is.null(opiClose()$err))
warning("opiClose() failed")
Response times to white-on-white Goldmann Size III targets for 12 subjects in dB units
Description
Response times to white-on-white Goldmann Size III targets for 12 subjects. The second column is the distance of the stimuli from measured threshold in HFA dB units. The threshold was determined by post-hoc fit of FoS curves to the data.
Usage
RtDbUnits
Format
An object of class data.frame with 30620 rows and 3 columns.
Details
A data frame with 30620 observations on the following 3 variables.
RtReaction time in ms.
DistDistance of stimuli from threshold in dB.
PersonIdentifier of each subject.
References
A.M. McKendrick, J. Denniss and A. Turpin. "Response times across the visual field: empirical observations and application to threshold determination". Vision Research 101 2014.
See Also
Response times to white-on-white Goldmann Size III targets for 12 subjects in sigma units
Description
Response times to white-on-white Goldmann Size III targets for 12 subjects. The second column is the distance of the stimuli from measured threshold in 'sigma' units. The threshold was determined by post-hoc fit of a cummulative gaussian FoS curve to the data for each location and subject. Sigma is the standard deviation of the fitted FoS.
Usage
RtSigmaUnits
Format
An object of class data.frame with 30620 rows and 3 columns.
Details
A data frame with 30620 observations on the following 3 variables.
RtReaction time in ms.
DistDistance of stimuli from threshold in sigma units.
PersonIdentifier of each subject.
References
A.M. McKendrick, J. Denniss and A. Turpin. "Response times across the visual field: empirical observations and application to threshold determination". Vision Research 101 2014.
See Also
ZEST
Description
An implementation of the Bayesian test procedures of King-Smith et al.
and Watson and Pelli. Note that we use the term pdf throughout as in the
original paper, even though they are discrete probability functions in this
implementation.
Usage
ZEST(
domain = 0:40,
prior = rep(1/length(domain), length(domain)),
likelihood = sapply(domain, function(tt) 0.03 + (1 - 0.03 - 0.03) * (1 -
stats::pnorm(domain, tt, 1))),
stopType = "S",
stopValue = 1.5,
minStimulus = utils::head(domain, 1),
maxStimulus = utils::tail(domain, 1),
maxSeenLimit = 2,
minNotSeenLimit = 2,
maxPresentations = 100,
minInterStimInterval = NA,
maxInterStimInterval = NA,
verbose = 0,
makeStim,
stimChoice = "mean",
...
)
ZEST.start(
domain = 0:40,
prior = rep(1/length(domain), length(domain)),
likelihood = sapply(domain, function(tt) 0.03 + (1 - 0.03 - 0.03) * (1 -
stats::pnorm(domain, tt, 1))),
stopType = "S",
stopValue = 1.5,
minStimulus = utils::head(domain, 1),
maxStimulus = utils::tail(domain, 1),
maxSeenLimit = 2,
minNotSeenLimit = 2,
maxPresentations = 100,
makeStim,
stimChoice = "mean",
...
)
ZEST.step(state, nextStim = NULL, fixedStimValue = NA, fixedResponse = NA)
ZEST.stop(state)
ZEST.final(state)
Arguments
domain |
Vector of values over which pdf is kept. |
prior |
Starting probability distribution over domain. Same length as |
likelihood |
Matrix where |
stopType |
|
stopValue |
Value for number of presentations ( |
minStimulus |
The smallest stimuli that will be presented. Could be different from
|
maxStimulus |
The largest stimuli that will be presented. Could be different from
|
maxSeenLimit |
Will terminate if |
minNotSeenLimit |
Will terminate if |
maxPresentations |
Maximum number of presentations regardless of |
minInterStimInterval |
If both |
maxInterStimInterval |
|
verbose |
|
makeStim |
A function that takes a dB value and numPresentations and returns an OPI datatype ready for passing to opiPresent. See examples. |
stimChoice |
A true ZEST procedure uses the |
... |
Extra parameters to pass to the opiPresent function |
state |
Current state of the ZEST returned by |
nextStim |
A valid object for |
fixedStimValue |
A number in |
fixedResponse |
Ignored if |
Details
This is an implementation of King-Smith et al.'s ZEST procedure and Watson and Pelli's QUEST procedure. All presentations are rounded to an element of the supplied domain.
Note this function will repeatedly call opiPresent for a stimulus until opiPresent
returns NULL (ie no error occurred).
The checkFixationOK function is called (if present in stim made from makeStim)
after each presentation, and if it returns FALSE, the pdf for that location is not changed
(ie the presentation is ignored), but the stim, number of presentations etc is recorded in
the state.
If more than one ZEST is to be interleaved (for example, testing multiple locations), then the
ZEST.start, ZEST.step, ZEST.stop and ZEST.final calls can maintain
the state of the ZEST after each presentation, and should be used. If only a single ZEST is
required, then the simpler ZEST can be used, which is a wrapper for the four functions
that maintain state. See examples below.
Value
Single location
ZEST returns a list containing
-
npresTotal number of presentations used. -
respSeqResponse sequence stored as a matrix: row 1 is dB values of stimuli, row 2 is 1/0 for seen/not-seen, row 3 is fixated 1/0 (always 1 ifcheckFixationOKnot present in stim objects returned frommakeStim). -
pdfsIfverboseis bigger than 0, then this is a list of the pdfs used for each presentation, otherwise NULL. -
finalThe mean/median/mode of the final pdf, depending onstimChoice, which is the determined threshold. -
opiRespA list of responses received from each successful call toopiPresentwithinZEST.
Multilple locations
ZEST.start returns a list that can be passed to ZEST.step, ZEST.stop, and
ZEST.final. It represents the state of a ZEST at a single location at a point in time
and contains the following.
-
nameZEST. -
pdfCurrent pdf: vector of probabilities the same length asdomain. -
numPresentationsThe number of timesZEST.stephas been called on this state. -
stimuliA vector containing the stimuli used at each call ofZEST.step. -
responsesA vector containing the responses received at each call ofZEST.step. -
responseTimesA vector containing the response times received at each call ofZEST.step. -
fixatedA vector containing TRUE/FALSE if fixation was OK according tocheckFixationOKfor each call ofZEST.step(defaults to TRUE ifcheckFixationOKnot present). -
opiRespA list of responses received from each call toopiPresentwithinZEST.step. A copy of all of the parameters supplied to ZEST.start:
domain,likelihood,stopType,stopValue,minStimulus,maxStimulus,maxSeenLimit,minNotSeenLimit,maxPresentations,makeStim,stimChoice,currSeenLimit,currNotSeenLimit, andopiParams.
ZEST.step returns a list containing
* state The new state after presenting a stimuli and getting a response.
* resp The return from the opiPresent call that was made.
ZEST.stop returns TRUE if the ZEST has reached its stopping criteria, and FALSE otherwise.
ZEST.final returns an estimate of threshold based on state. If state$stimChoice
is mean then the mean is returned. If state$stimChoice is mode then the
mode is returned. If state$stimChoice is median then the median is returned.
A list containing
* state The new state after presenting a stimuli and getting a response.
* resp The return from the opiPresent call that was made.
References
P.E. King-Smith, S.S. Grigsny, A.J. Vingrys, S.C. Benes, and A. Supowit. "Efficient and Unbiased Modifications of the QUEST Threshold Method: Theory, Simulations, Experimental Evaluation and Practical Implementation", Vision Research 34(7) 1994. Pages 885-912.
A.B. Watson and D.G. Pelli. "QUEST: A Bayesian adaptive psychophysical method", Perception and Psychophysics 33 1983. Pages 113-l20.
A. Turpin, P.H. Artes and A.M. McKendrick "The Open Perimetry Interface: An enabling tool for clinical visual psychophysics", Journal of Vision 12(11) 2012.
See Also
Examples
chooseOpi("SimHenson")
if(!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
##############################################
# This section is for single location ZESTs
##############################################
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}
repp <- function(...) sapply(1:50, function(i) ZEST(makeStim=makeStim, ...))
a <- repp(stopType="H", stopValue= 3, verbose=0, tt=30, fpr=0.03)
b <- repp(stopType="S", stopValue=1.5, verbose=0, tt=30, fpr=0.03)
c <- repp(stopType="S", stopValue=2.0, verbose=0, tt=30, fpr=0.03)
d <- repp(stopType="N", stopValue= 50, verbose=0, tt=30, fpr=0.03)
e <- repp(prior=dnorm(0:40,m=0,s=5), tt=30, fpr=0.03)
f <- repp(prior=dnorm(0:40,m=10,s=5), tt=30, fpr=0.03)
g <- repp(prior=dnorm(0:40,m=20,s=5), tt=30, fpr=0.03)
h <- repp(prior=dnorm(0:40,m=30,s=5), tt=30, fpr=0.03)
layout(matrix(1:2,1,2))
boxplot(lapply(list(a,b,c,d,e,f,g,h), function(x) unlist(x["final",])))
boxplot(lapply(list(a,b,c,d,e,f,g,h), function(x) unlist(x["npres",])))
##############################################
# This section is for multiple ZESTs
##############################################
makeStimHelper <- function(db,n, x, y) { # returns a function of (db,n)
ff <- function(db, n) db+n
body(ff) <- substitute({
s <- list(x=x, y=y, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
return(s)
}, list(x=x,y=y))
return(ff)
}
# List of (x, y, true threshold) triples
locations <- list(c(9,9,30), c(-9,-9,32), c(9,-9,31), c(-9,9,33))
# Setup starting states for each location
states <- lapply(locations, function(loc) {
ZEST.start(
domain=-5:45,
minStimulus=0,
maxStimulus=40,
makeStim=makeStimHelper(db,n,loc[1],loc[2]),
stopType="S", stopValue= 1.5, tt=loc[3], fpr=0.03, fnr=0.01)})
# Loop through until all states are "stop"
while(!all(st <- unlist(lapply(states, ZEST.stop)))) {
i <- which(!st) # choose a random,
i <- i[runif(1, min=1, max=length(i))] # unstopped state
r <- ZEST.step(states[[i]]) # step it
states[[i]] <- r$state # update the states
}
finals <- lapply(states, ZEST.final) # get final estimates of threshold
for(i in 1:length(locations)) {
#cat(sprintf("Location (%+2d,%+2d) ",locations[[i]][1], locations[[i]][2]))
#cat(sprintf("has threshold %4.2f\n", finals[[i]]))
}
if (!is.null(opiClose()$err))
warning("opiClose() failed")
Convert cd/\mbox{m}^2 to perimetric dB.
Description
Given a value in cd/\mbox{m}^2, return the equivalent dB value.
Default is to use HFA units, so maximum stimulus is 10000 apostilbs.
Usage
cdTodb(cd, maxStim = 10000/pi)
Arguments
cd |
Value to convert to dB in cd/ |
maxStim |
Stimulus value for 0dB in cd/ |
Value
A dB value for cd cd/\mbox{m}^2.
Examples
# candela to decibels
dB <- cdTodb(10000/pi) # 0 dB
dB <- cdTodb(1000/pi) # 10 dB
dB <- cdTodb(100/pi) # 20 dB
dB <- cdTodb(10/pi) # 30 dB
dB <- cdTodb(1/pi) # 40 dB
dB <- cdTodb(0.1/pi) # 50 dB
chooseOPI selects an OPI machine to use.
Description
It should be called before any other OPI functions.
Usage
chooseOPI(machine = NULL)
chooseOpi(machine = NULL)
Arguments
machine |
Machine name to use. Set to NULL to get a list. |
Value
NULL on success or list of machines otherwise.
Convert perimetric dB to cd/\mbox{m}^2
Description
Given a value in dB, return the cd/\mbox{m}^2
equivalent. Default is to use HFA units, so maximum stimulus is 10000
apostilbs.
Usage
dbTocd(db, maxStim = 10000/pi)
Arguments
db |
Value to convert to cd/ |
maxStim |
Stimulus value for 0dB in cd/ |
Value
cd/\mbox{m}^2 value for db dB.
Examples
# decibels to candela
cd <- dbTocd(0) # 10000/pi
cd <- dbTocd(10) # 1000/pi
cd <- dbTocd(20) # 100/pi
cd <- dbTocd(30) # 10/pi
cd <- dbTocd(40) # 1/pi
Convert degrees to pixels for machine 'machine'
Description
Convert degrees to pixels for machine 'machine'
Usage
degTopix(xy, machine = "compass")
Arguments
xy |
a 2 element vector c(x,y) where x and y are in pixels |
machine |
"compass" or ...? |
Value
xy converted to pixels (top-left is (0,0)) for the machine or
NA if machine is unknown
Examples
degTopix(c(0, 0), machine="compass") # c(960, 960) pixels
degTopix(c(-15, 2)) # c(495, 898) pixels
4-2 Staircase
Description
fourTwo is a 4-2 dB staircase beginning at level est
terminating after two reversals. The final estimate is the average of the
last two presentations. It also terminates if the minStimulus is
not seen twice, or the maxStimulus is seen twice.
Usage
fourTwo.start(est = 25, instRange = c(0, 40), verbose = FALSE, makeStim, ...)
fourTwo.step(state, nextStim = NULL)
fourTwo.stop(state)
fourTwo.final(state)
Arguments
est |
Starting estimate in dB |
instRange |
Dynamic range of the instrument c(min,max) in dB |
verbose |
True if you want each presentation printed |
makeStim |
A function that takes a dB value and numPresentations and returns an OPI datatype ready for passing to opiPresent |
... |
Extra parameters to pass to the opiPresent function |
state |
Current state of the fourTwo returned by
|
nextStim |
A valid object for |
Details
This is an implementation of a 4-2 1-up 1-down staircase. The
initial staircase starts at est and proceeds in steps of 4 dB until
the first reversal, and 2dB until the next reversal. The mean of the last
two presentations is taken as the threshold value. Note this function will
repeatedly call opiPresent for a stimulus until opiPresent
returns NULL (ie no error occurred). If more than one fourTwo is to
be interleaved (for example, testing multiple locations), then the
fourTwo.start, fourTwo.step, fourTwo.stop and
fourTwo.final calls can maintain the state of the fourTwo after
each presentation, and should be used. See examples below.
Value
Multiple locations
fourTwo.start returns a list that can be passed to fourTwo.step,
fourTwo.stop, and fourTwo.final. It represents the state of a fourTwo
at a single location at a point in time and contains the following.
-
namefourTwo A copy of all of the parameters supplied to fourTwo.start:
startingEstimate=est,minStimulus=instRange[1],maxStimulus=instRange[2],makeStim, andopiParams=list(...)-
currentLevelThe next stimulus to present. -
lastSeenThe last seen stimulus. -
lastResponseThe last response given. -
stairResultThe final result if finished (initiallyNA). -
finished"Not"if staircase has not finished, or one of"Rev"(finished due to 2 reversals),"Max"(finished due to 2maxStimulusseen),"Min"(finished due to 2minStimulusnot seen) -
numberOfReversalsNumber of reversals so far. -
currSeenLimitNumber of timesmaxStimulushas been seen. -
currNotSeenLimitNumber of timesminStimulusnot seen. -
numPresentationsNumber of presentations so far. -
stimuliVector of stimuli shown at each call tofourTwo.step. -
responsesVector of responses received (1 seen, 0 not) received at each call tofourTwo.step. -
responseTimesVector of response times received at each call tofourTwo.step.
fourTwo.step returns a list containing
-
stateThe new state after presenting a stimuli and getting a response. -
respThe return from theopiPresentcall that was made.
fourTwo.stop returns TRUE if the staircase is finished (2 reversals, or maxStimulus
is seen twice or minStimulus is not seen twice).
fourTwo.final returns the final estimate of threshold (mean of last
two reversals). This issues a warning if called before the staircase has
finished, but still returns a value.
See Also
Examples
# Stimulus is Size III white-on-white as in the HFA
makeStim <- function(db, n) {
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500)
class(s) <- "opiStaticStimulus"
return(s)
}
chooseOpi("SimHenson")
if (!is.null(opiInitialize(type="C", cap=6)$err))
stop("opiInitialize failed")
##############################################
# This section is for multiple fourTwos
##############################################
makeStimHelper <- function(db,n, x, y) { # returns a function of (db,n)
ff <- function(db, n) db+n
body(ff) <- substitute({
s <- list(x=x, y=y, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500)
class(s) <- "opiStaticStimulus"
return(s)}, list(x=x,y=y))
return(ff)
}
# List of (x, y, true threshold) triples
locations <- list(c(9,9,30), c(-9,-9,32), c(9,-9,31), c(-9,9,33))
# Setup starting states for each location
states <- lapply(locations, function(loc) {
fourTwo.start(makeStim=makeStimHelper(db,n,loc[1],loc[2]),
tt=loc[3], fpr=0.03, fnr=0.01)})
# Loop through until all states are "stop"
while(!all(st <- unlist(lapply(states, fourTwo.stop)))) {
i <- which(!st) # choose a random,
i <- i[runif(1, min=1, max=length(i))] # unstopped state
r <- fourTwo.step(states[[i]]) # step it
states[[i]] <- r$state # update the states
}
finals <- lapply(states, fourTwo.final) # get final estimates of threshold
for(i in 1:length(locations)) {
cat(sprintf("Location (%+2d,%+2d) ",locations[[i]][1], locations[[i]][2]))
cat(sprintf("has threshold %4.2f\n", finals[[i]]))
}
if (!is.null(opiClose()$err))
warning("opiClose() failed")
Present kinetic on Kowa AP7000 (internal use)
Description
Implementation of opiPresent for the Kowa AP7000 machine. Version for opiKineticStimulus.
This is for internal use only. Use opiPresent() with
stim as class opiStaticStimulus and you will get the Value back.
Arguments
stim |
Stimulus to present (a list, see details). |
Details
stim is a list containing at least the following 3 elements:
-
path, a list containingx2 x-coordinate in degrees (floating point) (range $[-80,80]$) andy, list of 2 y-coordinate in degrees (floating point) (range $[-70,65]$). -
sizeslist of 1 size; one of.opi_env$KowaAP7000$SIZES_DEGREES. -
colorslist of 1 color; one of.opi_env$KowaAP7000$COLOR_WHITE,.opi_env$KowaAP7000$COLOR_GREEN,.opi_env$KowaAP7000$COLOR_BLUE, or.opi_env$KowaAP7000$COLOR_RED -
levelslist of 1 level; luminance in cd/\mbox{m}^2, and is rounded to the nearest whole dB for display (range 0 to 50). 0dB is 10000aps. -
speedslist of 1 speed; degrees per second range $[3, 5]$. -
durationof stimulus on in milliseconds (range $[100, 1200]$). -
responseWindowfrom start of stimulus presentation in milliseconds (max 5000).
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. -
timeReaction time (if seen). -
xLocation of button press in degrees. -
yLocation of button press in degrees.
Present static on Kowa AP7000 (internal use)
Description
Implementation of opiPresent for the Kowa AP7000 machine. Version for opiStaticStimulus.
This is for internal use only. Use opiPresent() with
stim as class opiStaticStimulus and you will get the Value back.
Arguments
stim |
Stimulus to present (a list, see details). |
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
Details
stim is a list containing at least the following 3 elements:
-
x, x-coordinate in degrees (floating point) (range $[-80,80]$). -
y, y-coordinate in degrees (floating point) (range $[-70,65]$). -
levelis luminance in cd/\mbox{m}^2, and is rounded to the nearest whole dB for display (range 0 to 50). 0dB is 10000aps. -
durationof stimulus on in milliseconds (range $[100, 1200]$). -
responseWindowfrom start of stimulus presentation in milliseconds (max 5000). -
sizeone of.opi_env$KowaAP7000$SIZES_DEGREES. -
colorone of.opi_env$KowaAP7000$COLOR_WHITE,.opi_env$KowaAP7000$COLOR_GREEN,.opi_env$KowaAP7000$COLOR_BLUE, or.opi_env$KowaAP7000$COLOR_RED
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. -
timeReaction time (if seen). -
pupilX -
pupilY -
purkinjeX -
purkinjeY
Present temporal stim on Kowa AP7000 (internal use)
Description
Implementation of opiPresent for the Kowa AP7000 machine. Version for opiKineticStimulus.
This is for internal use only. Use opiPresent() with
stim as class opiStaticStimulus and you will get the Value back.
Present Kinetic stimuli on O900 (internal use)
Description
Implementation of opiPresent for Kinetic stimuli on the Octopus090 machine.
This is for internal use only. Use opiPresent() with
these Arguments and stim as class opiKineticStimulus
and you will get the Value back.
Arguments
stim |
Stimulus to present (a list, see details). |
nextStim |
Ignored. |
Details
stim is a list containing at least the following 3 elements:
-
path, A list of $(x,y)$ coordinates in degrees that is usable byxy.coords(). -
sizes, A list wheresizes[i]is the size of stimulus (diameter in degrees) to use for the section of path specified bypath[i]..path[i+1]. Rounded to nearest Goldmann size. -
levelsA list wherelevels[i]is the stimulus luminance in cd/\mbox{m}^2to use for the section of path specified bypath[i]..path[i+1]. -
speedsA list wherespeeds[i]is the speed in degrees per second to use for the section of path specified bypath[i]..path[i+1].
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. -
timeReaction time (if seen). -
xCoordinate where button was pressed (degrees - i guess). -
yCoordinate where button was pressed (degrees - i guess).
Present static on O900 (internal use)
Description
Implementation of opiPresent for the Octopus090 machine. Version for opiStaticStimulus.
This is for internal use only. Use opiPresent() with
stim as class opiStaticStimulus and you will get the Value back.
Arguments
stim |
Stimulus to present (a list, see details). |
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
F310 |
If |
Details
stim is a list containing at least the following 3 elements:
-
x, x-coordinate in degrees (floating point) (range $[-30,30]$). -
y, y-coordinate in degrees (floating point) (range $[-30,30]$). -
levelis luminance in cd/\mbox{m}^2, and is rounded to the nearest whole dB for display (range 0 to 50). 0dB is 10000aps.
It can also contain:
-
responseWindowfrom start of stimulus presentation in milliseconds (default is 1500). -
durationof stimulus on in milliseconds (default 200). -
colorone of.opi_env$O900$STIM_WHITE,.opi_env$O900$STIM_BLUEor.opi_env$O900$STIM_RED. It must be same as that initialised byopiSetup()oropiInitialize()(default.opi_env$O900$STIM_WHITE). -
sizeof stimulus diameter in degrees (default Size III == 0.43). This is rounded to the nearest support Goldmann size.
If responses are taken from the F310 Controller then
If the L button is pressed,
seenis set to 1.If the R button is pressed,
seenis set to 2.If no button is pressed within
responseWindow, thenseenis set to 0.
If stim is null, always return err = NULL status.
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. (See details for F310) -
timeReaction time (if seen).
Present Temporal stimuli on O900 (internal use)
Description
Implementation of opiPresent for Temporal stimuli on the Octopus090 machine.
This is for internal use only. Use opiPresent() with
these Arguments (stim as class opiTemporalStimulus)
and you will get the Value back.
Arguments
stim |
Stimulus to present (a list, see details). |
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
Details
stim is a list containing at least the following 3 elements:
-
x, x-coordinate in degrees (floating point) (range $[-30,30]$). -
y, y-coordinate in degrees (floating point) (range $[-30,30]$). -
rateis frequency in Hz.
It can also contain:
-
responseWindowfrom start of stimulus presentation in milliseconds (default is 1500). -
durationof stimulus on in milliseconds (default 200). -
sizeof stimulus diameter in degrees (default Size III == 0.43). This is rounded to the nearest support Goldmann size.
If stim is null, always return err = NULL status.
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. -
timeReaction time (if seen).
Open a socket on ip and port.
Description
Internal use only.
Usage
open_socket(ip, port, machineName)
Arguments
ip |
IP address of socket |
port |
TCP port of socket |
machineName |
Machine name for error message |
Value
Socket or NULL on error
Calls opiClose_for_MACHINE as appropriate.
Description
Specific parameters and return values can be seen in the machine specific versions listed below in the ’See Also’.
Usage
opiClose()
Value
Each implementation should(!) return a list with at least the following elements:
-
errNULLif no error, otherwise a string describing the error.
See Also
opiClose_for_Compass(),
opiClose_for_Octopus900(),
opiClose_for_ImoVifa(),
opiClose_for_PhoneHMD(), opiClose_for_Display(), opiClose_for_PicoVR(),
opiClose_for_SimNo(), opiClose_for_SimYes(), opiClose_for_SimHenson(),
opiClose_for_SimHensonRT(),
opiClose_for_SimGaussian()
Implementation of opiClose for the Compass machine.
Description
This is for internal use only. Use opiClose() with
these Arguments and you will get the Value back.
Value
A list with elements
-
err, which is an error code, NULL for no error -
fixations, which is a matrix one row per fixation and three columns:-
time(same astime_hwinopiPresent) -
x(degrees relative to the centre of the image returned byopiInitialise- not the PRL) -
y(as for x)
-
Implementation of opiClose for the Display machine.
Description
This is for internal use only. Use opiClose() after
chooseOPI("Display") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("Display")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiClose()
## End(Not run)
Implementation of opiClose for the ImoVifa machine.
Description
This is for internal use only. Use opiClose() after
chooseOPI("ImoVifa") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("ImoVifa")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiClose()
## End(Not run)
Implementation of opiClose for the Kowa AP7000 machine.
Description
This is for internal use only. Use opiClose() with the same parameters.
Value
Returns list(err = NULL).
Implementation of opiClose for the MAIA machine.
Description
This is for internal use only. Use opiClose() with
these Arguments and you will get the Value back.
Value
A list with elements
-
err, which is an error code, NULL for no error -
fixations, which is a matrix one row per fixation and four columns:-
time(same astime_hwinopiPresent) -
flagsBit 0(same astime_hwinopiPresent)bit 0: 1 if the tracking event is valid (quality>threshold), 0 otherwise (in this case, x and y shall be =0)
bit 2: 1 if a stimulus was being presented during the tracking event, 0 otherwise
bit 3: 1 if the tracking was enabled during the tracking event, 0 otherwise
bits 16..31: ID of the stimulus being presented (if bit 2 == 1)
-
x(degrees relative to the centre of the image returned byopiInitialise- not the PRL) -
y(as for x)
-
Implementation of opiClose for the O600 machine.
Description
This is for internal use only. Use opiClose() with the same parameters.
Value
Returns list(err = NULL).
Implementation of opiClose for the Octopus900 machine.
Description
This is for internal use only. Use opiClose() with the same parameters.
Value
Returns list(err = NULL).
Implementation of opiClose for the PhoneHMD machine.
Description
This is for internal use only. Use opiClose() after
chooseOPI("PhoneHMD") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PhoneHMD")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiClose()
## End(Not run)
Implementation of opiClose for the PicoVR machine.
Description
This is for internal use only. Use opiClose() after
chooseOPI("PicoVR") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PicoVR")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiClose()
## End(Not run)
opiClose_for_SimGaussian
Description
Does nothing.
Value
A list with elements:
-
errorAlwaysFALSE. -
msgA string "Close OK".
opiClose_for_SimHenson
Description
Does nothing.
Value
A list with elements:
-
errAlwaysNULL.
opiClose_for_SimHensonRT
Description
Does nothing.
Value
A list with elements:
-
errAlwaysNULL.
opiClose_for_SimNo
Description
Does nothing.
Value
A list with elements:
-
errorAlwaysFALSE. -
msgA string "Close OK".
opiClose_for_SimYes
Description
Does nothing.
Value
A list with elements:
-
errAlwaysNULL.
Calls opiInitialise_for_MACHINE as appropriate.
Description
Establishes connection with the device and a Monitor (aka Server) if appropriate. Sends any startup parameters that might be needed by the machine. Specific parameters and return values can be seen in the machine specific versions listed below in the ’See Also’.
Usage
opiInitialise(...)
opiInitialize(...)
Arguments
... |
Parameters specific to each machine as described in the 'See Also' functions. |
Value
A list containing at least the following elements:
-
errNULLif no error, otherwise a string describing the error.
See Also
opiInitialise_for_ImoVifa(),
opiInitialise_for_PhoneHMD(), opiInitialise_for_Display(), opiInitialise_for_PicoVR(),
opiInitialise_for_Octopus900(),
opiInitialise_for_Compass(),
opiInitialise_for_SimNo(), opiInitialise_for_SimYes(), opiInitialise_for_SimHenson(),
opiInitialise_for_SimHensonRT(),
opiInitialise_for_SimGaussian()
Implementation of opiInitialise for the ImoVifa machine.
Description
This is for internal use only. Use opiInitialise() with
these Arguments and you will get the Value back.
Arguments
ip |
IP address on which server is listening as a string |
port |
Port number on which server is listening |
... |
Could be used for fake compass, simulations, etc |
Details
Warning: this returns a list, not a single error code.
Value
A list with elements:
-
errNULL if successful, not otherwise. -
prlA pair giving the (x,y) in degrees of the Preferred Retinal Locus detected in the initial alignment. -
onha pair giving the (x,y) in degrees of the ONH as selected by the user. -
imageraw bytes being the JPEG compressed infra-red image acquired during alignment.
Examples
## Not run:
# Set up the Compass
chooseOpi("Compass")
result <- opiInitialize(ip = "192.168.1.7", port = 44965)
if (is.null(result$err))
print(result$prl)
## End(Not run)
Implementation of opiInitialise for the Display machine.
Description
This is for internal use only. Use opiInitialise() after
chooseOPI("Display") to call this function.
Arguments
address |
A list containing:
|
Details
port can take on values in the range [0, 65535].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("Display")
result <- opiInitialise(address = list(port = 50001, ip = "localhost"))
## End(Not run)
Implementation of opiInitialise for the ImoVifa machine.
Description
This is for internal use only. Use opiInitialise() after
chooseOPI("ImoVifa") to call this function.
Arguments
address |
A list containing:
|
Details
port can take on values in the range [0, 65535].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("ImoVifa")
result <- opiInitialise(address = list(port = 50001, ip = "localhost"))
## End(Not run)
Implementation of opiInitialise for the Octopus900 machine.
Description
This is for internal use only. Use opiInitialise() with
these Arguments and you will get the Value back.
Arguments
ip |
IP address of AP7000 machine (as string) |
port |
Port number on which AP7000 server is listening |
Details
If the chosen OPI implementation is KowaAP7000, then you must specify
the IP address and port of the AP-7000 server.
Value
list(err = NULL)
Examples
## Not run:
# Set up the Kowa AP-7000
chooseOpi("KowaAP7000")
opiInitialize(ip="192.168.1.7", port=44965)
## End(Not run)
Implementation of opiInitialise for the ImoVifa machine.
Description
This is for internal use only. Use opiInitialise() with
these Arguments and you will get the Value back.
Arguments
ip |
IP address on which server is listening as a string |
port |
Port number on which server is listening |
... |
Could be used for fake MAIA, simulations, etc |
Details
Establishes socket connection but sends not messages (see opiSetup_for_MAIA).
Value
A list with elements:
-
errNULL if successful, not otherwise.
Examples
## Not run:
# Set up the MAIA
chooseOpi("MAIA")
result <- opiInitialize(ip = "192.168.1.7", port = 5555)
if (is.null(result$err))
print(result$err)
## End(Not run)
Implementation of opiInitialise for the O600 machine.
Description
This is for internal use only. Use opiInitialise() with
these Arguments and you will get the Value back.
Arguments
ipAddress |
The IP address of the O600 as a string. |
eye |
Either "left" or "right". |
pupilTracking |
TRUE to turn on IR illumination and set pupil black level (which happens at the first stimulus presentation). |
pulsar |
TRUE for pulsar stimulus, FALSE for size III white-on-white. |
eyeControl |
One of * 0 is off * 1 is eye blink * 2 is eye blink, forehead rest, fixation control * 3 is eye blink, forehead rest, fixation control, fast eye movements |
Details
The default background and stimulus setup is to white-on-white perimetry.
Uses port 50000 on the O600.
Examples
## Not run:
# Set up the O600
chooseOpi("O600")
opiInitialize(ip="192.168.1.7", eye = "left")
## End(Not run)
Implementation of opiInitialise for the Octopus900 machine.
Description
This is for internal use only. Use opiInitialise() with
these Arguments and you will get the Value back.
Arguments
serverPort |
port number on which server is listening for "Octopus900" |
eyeSuiteSettingsLocation |
dir name containing EyeSuite settings for "Octopus900" |
eye |
eye; "right" or "left" for "Octopus900", "Octopus600" |
gazeFeed |
NA or a folder name for "Octopus900" |
bigWheel |
FALSE (standard machine), TRUE for modified aperture wheel for "Octopus900" |
pres_buzzer |
0 (no buzzer),1, 2, 3 (max volume) for "Octopus900" |
resp_buzzer |
0 (no buzzer),1, 2, 3 (max volume) for "Octopus900" |
zero_dB_is_10000_asb |
Is 0 dB 10000 apostilb (TRUE) or or 4000 (FALSE) for "Octopus900" |
Details
If the chosen OPI implementation is Octopus900, then you must specify
a directory and the eye to be tested.
serverPort is the TCP/IP port on which the server is listening (on
localhost).
eyeSuiteSettingsLocation is the folder name containing the EyeSuite
setting files, and should include the trailing slash.
eye must be either "left" or "right".
gazeFeed is the name of an existing folder into which the video frames
of eye tracker are recorded. Set to NA for no recording.
bigWheel is FALSE for a standard Octopus 900 machine. Some
research machines are fitted with an alternate aperture wheel that has 24
sizes, which are accessed with bigWheel is TRUE. The mapping
from size to 'hole on wheel' is hard coded; see code for details.
If pres_buzzer is greater than zero, a buzzer will sound with each
stimuli presented.
If resp_buzzer is greater than zero, a buzzer will sound with each
button press (response). The volume can be one of 0 (no buzzer), 1, 2, or 3
(max volume). If both buzzers are more than zero, the maximum of the two will
be used as the volume.
If zero_dB_is_10000_asb is TRUE then 0 dB is taken as 10000
apostilbs, otherwise 0 dB is taken as 4000 apostilbs.
Value
A list containing err which is
NULL if successful
1 if Octopus900 is already initialised by a previous call to
opiInitialize2 if some error occurred that prevented initialisation.
Examples
## Not run:
chooseOpi("Octopus900")
res <- opiInitialize(serverPort = 50001,
eyeSuiteSettingsLocation = "C:/ProgramData/Haag-Streit/EyeSuite/",
eye = "", gazeFeed = "", bigWheel = FALSE,
pres_buzzer = 0, resp_buzzer = 0, zero_dB_is_10000_asb = TRUE)
if (!is.null(res$err))
stop("opiInitialize failed")
## End(Not run)
Implementation of opiInitialise for the PhoneHMD machine.
Description
This is for internal use only. Use opiInitialise() after
chooseOPI("PhoneHMD") to call this function.
Arguments
address |
A list containing:
|
Details
port can take on values in the range [0, 65535].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PhoneHMD")
result <- opiInitialise(address = list(port = 50001, ip = "localhost"))
## End(Not run)
Implementation of opiInitialise for the PicoVR machine.
Description
This is for internal use only. Use opiInitialise() after
chooseOPI("PicoVR") to call this function.
Arguments
address |
A list containing:
|
Details
port can take on values in the range [0, 65535].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PicoVR")
result <- opiInitialise(address = list(port = 50001, ip = "localhost"))
## End(Not run)
opiInitialize_for_SimGaussian
Description
Simulates responses using a Frequency of Seeing (FoS) curve.
The FoS is modelled as a cumulative Gaussian function with standard deviation
equal to sd as provided and the mean as the true threshold given as tt
opiPresent.
All values are in dB relative to maxStim.
This is for internal use only, use opiInitialize().
Arguments
sd |
Standard deviation of Cumulative Gaussian. |
maxStim |
The maximum stimuls value (0 dB) in cd/ |
... |
Any other parameters you like, they are ignored. |
Value
A list with elements:
-
errNULL if initialised, a message otherwise
Examples
# Set up a simple simulation for white-on-white perimetry
chooseOpi("SimGaussian")
res <- opiInitialize(sd = 2.5)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$msg))
opiInitialise_for_SimHenson
Description
Simulates responses using a Frequency of Seeing (FoS) curve.
For internal use only, use opiInitialize().
The FoS is modelled as a cumulative Gaussian function over dB with
standard deviation equal to min(cap, exp( A * t + B)), where
t is the threshold/mean of the FoS in dB.
All values are in dB relative to maxStim.
Arguments
type |
A single character that is:
|
A |
Coefficient of |
B |
Addend of |
cap |
Maximum dB value for the stdev of the FoS curve. |
maxStim |
The maximum stimulus value (0 dB) in cd/ |
... |
Any other parameters you like, they are ignored. |
Value
A list with elements:
-
errNULL if initialised, msg otherwise
Examples
# Set up a simple simulation for white-on-white perimetry
chooseOpi("SimHenson")
res <- opiInitialize(type = "C", cap = 6)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$err))
opiInitialise_for_SimHensonRT
Description
Simulates responses using a Frequency of Seeing (FoS) curve.
For internal use only, use opiInitialize().
The FoS is modelled as a cumulative Gaussian function over dB with
standard deviation equal to min(cap, exp( A * t + B)), where
t is the threshold/mean of the FoS in dB.
All values are in dB relative to maxStim.
Arguments
type |
A single character that is:
|
A |
Coefficient of |
B |
Addend of |
cap |
Maximum dB value for the stdev of the FoS curve. |
maxStim |
The maximum stimulus value (0 dB) in cd/ |
rtData |
A data.frame with colnames == "Rt", "Dist", "Person" (or NULL for default). |
rtFP |
A response time for false positives ??? for "SimHensonRT" |
... |
Any other parameters you like, they are ignored. |
Details
If the chosen OPI implementation is SimHensonRT, then the first six
parameters are as in SimHenson, and rtData is a data frame
with at least 2 columns: "Rt", response time; and "Dist",
signifying that distance between assumed threshold and stimulus value in
your units.
This package contains RtSigmaUnits or RtDbUnits that can be
loaded with the commands data(RtSigmaUnits) or data(RtDbUnits),
and are suitable to pass as values for rtData.
rtFp gives the vector of values in milliseconds from which a response
time for a false positive response is randomly sampled.
Value
A list with elements:
-
errNULL if initialised, string msg otherwise
Examples
# Set up a simple simulation for white-on-white perimetry
# and display the stimuli in a plot region and simulate response times
chooseOpi("SimHensonRT")
data(RtSigmaUnits)
oi <- opiInitialize(type="C", cap=6, display=NA, rtData=RtSigmaUnits, rtFP=1:100)
if (!is.null(oi$err))
stop("opiInitialize failed")
# Set up a simple simulation for white-on-white perimetry
chooseOpi("SimHenson")
res <- opiInitialize(type = "C", cap = 6)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$err))
opiInitialise_for_SimNo
Description
Does nothing.
Arguments
... |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
opiInitialise_for_SimYes
Description
Does nothing.
Arguments
... |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Description
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Usage
opiKineticStimulus()
Calls opiPresent_for_MACHINE as appropriate.
Description
Specific parameters and return values can be seen in the machine specific versions listed below in the ’See Also’.
Usage
opiPresent(stim, ...)
Arguments
stim |
A stimulus object or list as described for each machine in the 'See Also' methods. |
... |
Other arguments that might be needed by each machine in the 'See Also' methods. |
Value
Each implementation should(!) return a list with at least the following elements:
-
errNULLif no error, otherwise a string describing the error. -
seenTRUEif stimulus seen,FALSEotherwise -
timeResponse time from onset of stimulus in milliseconds.
See Also
opiPresent_for_Compass(),
opiPresent_for_Octopus900(),
opiPresent_for_ImoVifa(),
opiPresent_for_PhoneHMD(), opiPresent_for_Display(), opiPresent_for_PicoVR(),
opiPresent_for_SimNo(), opiPresent_for_SimYes(), opiPresent_for_SimHenson(),
opiPresent_for_SimHensonRT(),
opiPresent_for_SimGaussian()
Implementation of opiPresent for the Compass machine.
Description
This is for internal use only. Use opiSetup() with
these Arguments and you will get the Value back.
Arguments
stim |
A list of stimulus parameters (see Details). |
nextStim |
Unused - included for compliance with OPI standard. |
Details
If the chosen OPI implementation is Compass, then nextStim
is ignored. Note that the dB level is rounded to the nearest integer.
If tracking is on, then this will block until the tracking is obtained, and the stimulus presented.
stim is a list containing some or all of the following elements:
-
x, x-coordinate in degrees (floating point) (range $[-30,30]$). -
y, y-coordinate in degrees (floating point) (range $[-30,30]$). -
levelis luminance in cd/\mbox{m}^2, and is rounded to the nearest whole dB for display (range 0 to 50). 0dB is 10000aps. -
responseWindowis in milliseconds (range 0 to 2680).
Stimulus duration is assumed to be 200ms, and size is assumed to be Goldmann III (0.43 degrees diameter), color is assumed to be white. These cannot be changed.
Value
A list containing:
-
err0 all clear, >= 1 some error codes (eg cannot track, etc) (integer) -
seenFALSEfor not seen,TRUEfor seen (button pressed in response window) -
timeResponse time in ms (integer) since stimulus onset, -1 for not seen -
time_recTime since epoch when command was received at Compass (integer ms) -
time_presTime since epoch that stimulus was presented (integer ms) -
num_track_eventsNumber of tracking events that occurred during presentation (integer) -
num_motor_failsNumber of times motor could not follow fixation movement during presentation (integer) -
pupil_diamPupil diameter in mm (float) -
loc_xPixels integer, location in image of presentation (integer) -
loc_yPixels integer, location in image of presentation (integer)
Examples
## Not run:
# Set up the Compass
chooseOpi("Compass")
result <- opiInitialize(ip = "192.168.1.7", port = 44965)
if (!is.null(result$err))
stop("Initialisation failed")
#' @param x X location of stim in degrees
#' @param y Y location of stim in degrees
#' @param size If 3, Goldmann III, else V
#' @param db Value in dB
#' @return stim object ready for opiPresent
makeStim <- function(x, y, size, db) {
s <- list(x = x, y = y, level = dbTocd(db, 10000 / pi),
size = ifelse(size == 3, 0.43, 1.77),
duration = 200, responseWindow = 1500)
class(s) <- "opiStaticStimulus"
return(s)
}
result <- opiPresent(makeStim(9, 9, 3, 10))
## End(Not run)
Implementation of opiPresent for the Display machine.
Description
This is for internal use only. Use opiPresent() after
chooseOPI("Display") to call this function.
Arguments
stim |
A list containing:
|
... |
Parameters for other opiPresent implementations that are ignored here. |
Details
Elements in lum can take on values in the range [0.0, 1.0E10].
stim.length can take on values in the range [1, 2147483647].
Elements in color1 can take on values in the range [0.0, 1.0].
Elements in sx can take on values in the range [0.0, 180.0].
Elements in sy can take on values in the range [0.0, 180.0].
Elements in eye can take on values in the set
{"left", "right", "both", "none"}.
Elements in t can take on values in the range [0.0, 1.0E10].
w can take on values in the range [0.0, 1.0E10].
Elements in x can take on values in the range [-90.0, 90.0].
Elements in y can take on values in the range [-90.0, 90.0].
Elements in envSdx can take on values in the range
[-1.0E10, 1.0E10].
Elements in envSdy can take on values in the range
[-1.0E10, 1.0E10].
Elements in envRotation can take on values in
the range [-1.0E10, 1.0E10].
Elements in type can take on values in the set
{"flat", "checkerboard", "sine", "squaresine", "g1",
"g2", "g3", "text", "image"}.
Elements in frequency can take on values in the
range [0.0, 300.0].
Elements in color2 can take on values in the range [0.0, 1.0].
Elements in fullFoV can take on values in the
range [-1.0E10, 1.0E10].
Elements in phase can take on values in the range [0.0, 1.0E10].
Elements in shape can take on values in the set
{"triangle", "square", "polygon", "hollow_triangle",
"hollow_square", "hollow_polygon", "cross", "maltese",
"circle", "annulus", "optotype", "text", "model"}.
Elements in rotation can take on values in the range [0.0, 360.0].
Elements in texRotation can take on values in
the range [0.0, 360.0].
Elements in defocus can take on values in the range [0.0, 1.0E10].
Elements in envType can take on values in the set
{"none", "square", "circle", "gaussian"}.
Elements in contrast can take on values in the range [0.0, 1.0].
Elements in optotype can take on values in the
set {"a", "b", "c", "d", "e", "f", "g", "h", "i",
"j", "k", "l", "m", "n", "o", "p", "q", "r", "s", "t",
"u", "v", "w", "x", "y", "z"}.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error. -
timeResponse time from stimulus onset if button pressed (ms). -
seen'1' if seen, '0' if not.
See Also
Examples
## Not run:
chooseOpi("Display")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiPresent(stim = list(lum = list(300.0), stim.length = 1, color1 = list(list(1.0,
1.0, 1.0)), sx = list(1.72), sy = list(1.72),
eye = list("LEFT"), t = list(200.0), w = 1500.0, x = list(0.0), y = list(0.0)))
## End(Not run)
Implementation of opiPresent for the ImoVifa machine.
Description
This is for internal use only. Use opiPresent() after
chooseOPI("ImoVifa") to call this function.
Arguments
stim |
A list containing:
|
... |
Parameters for other opiPresent implementations that are ignored here. |
Details
Elements in lum can take on values in the range [0.0, 1.0E10].
stim.length can take on values in the range [1, 2147483647].
Elements in color1 can take on values in the range [0.0, 1.0].
Elements in sx can take on values in the range [0.0, 180.0].
Elements in sy can take on values in the range [0.0, 180.0].
Elements in eye can take on values in the set
{"left", "right", "both", "none"}.
Elements in t can take on values in the range [0.0, 1.0E10].
w can take on values in the range [0.0, 1.0E10].
Elements in x can take on values in the range [-90.0, 90.0].
Elements in y can take on values in the range [-90.0, 90.0].
Elements in envSdx can take on values in the range
[-1.0E10, 1.0E10].
Elements in envSdy can take on values in the range
[-1.0E10, 1.0E10].
Elements in envRotation can take on values in
the range [-1.0E10, 1.0E10].
Elements in type can take on values in the set
{"flat", "checkerboard", "sine", "squaresine", "g1",
"g2", "g3", "text", "image"}.
Elements in frequency can take on values in the
range [0.0, 300.0].
Elements in color2 can take on values in the range [0.0, 1.0].
Elements in fullFoV can take on values in the
range [-1.0E10, 1.0E10].
Elements in phase can take on values in the range [0.0, 1.0E10].
Elements in shape can take on values in the set
{"triangle", "square", "polygon", "hollow_triangle",
"hollow_square", "hollow_polygon", "cross", "maltese",
"circle", "annulus", "optotype", "text", "model"}.
Elements in rotation can take on values in the range [0.0, 360.0].
Elements in texRotation can take on values in
the range [0.0, 360.0].
Elements in defocus can take on values in the range [0.0, 1.0E10].
Elements in envType can take on values in the set
{"none", "square", "circle", "gaussian"}.
Elements in contrast can take on values in the range [0.0, 1.0].
Elements in optotype can take on values in the
set {"a", "b", "c", "d", "e", "f", "g", "h", "i",
"j", "k", "l", "m", "n", "o", "p", "q", "r", "s", "t",
"u", "v", "w", "x", "y", "z"}.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error. -
eyedStartDiameter of pupil at stimulus onset (pixels). -
eyexEndx co-ordinate of pupil at button press (pixels from image centre. Image is 640x480, left < 0). Note that for multi-part stimuli (t=0), the eye taken is the last eye in the list of components. Not valid for not-seen response. -
eyeyEndy co-ordinate (pixels). See eyexEnd for more details. (up > 0). Not valid for not-seen response. -
eyedEndDiameter of pupil at button press or response window expiry (pixels). Not valid for not-seen response. -
eyexStartx co-ordinates of pupil at stimulus onset (pixels from image centre. Image is 640x480, left < 0). For a multi-part stimulus (t=0), the eye taken is the first eye in the list of components. -
timeResponse time from stimulus onset if button pressed (ms). -
seen'1' if seen, '0' if not. -
eyeyStarty co-ordinates (pixels from image centre). See eyexStart for more details. (up > 0)
See Also
Examples
## Not run:
chooseOpi("ImoVifa")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiPresent(stim = list(lum = list(300.0), stim.length = 1, color1 = list(list(1.0,
1.0, 1.0)), sx = list(1.72), sy = list(1.72),
eye = list("LEFT"), t = list(200.0), w = 1500.0, x = list(0.0), y = list(0.0)))
## End(Not run)
Implementation of opiPresent for the KowaAP7000 machine.
Description
This is for internal use only. Use opiPresent() with the same arguments and
the class of stim as one of opiStaticStimulus, opiTemporalStimulus, or opiKineticStimulus.
Arguments
stim |
Stimulus to present (a list, see |
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
Value
See kowa.presentStatic, kowa.presentTemporal, or kowa.presentKinetic.
See Also
kowa.presentStatic, kowa.presentTemporal, kowa.presentKinetic
Examples
## Not run:
chooseOpi("KowaAP7000")
if (!is.null(opiInitialize()$err))
stop("opiInitialize failed")
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
print(opiPresent(s, NULL))
## End(Not run)
Implementation of opiPresent for the MAIA machine.
Description
This is for internal use only. Use opiSetup() with
these Arguments and you will get the Value back.
Arguments
stim |
A list of stimulus parameters (see Details). |
nextStim |
Unused - included for compliance with OPI standard. |
Details
If the chosen OPI implementation is MAIA, then nextStim
is ignored. Note that the dB level is rounded to the nearest integer.
If tracking is on, then this will block until the tracking is obtained, and the stimulus presented.
stim is a list containing some or all of the following elements:
-
x, x-coordinate in degrees (floating point) (range $[-30,30]$). -
y, y-coordinate in degrees (floating point) (range $[-30,30]$). -
levelis luminance in cd/\mbox{m}^2, and is rounded to the nearest whole dB for display (range 0 to 50). 0dB is 10000aps. -
responseWindowis in milliseconds (range 0 to 2680).
Stimulus duration is assumed to be 200ms, and size is assumed to be Goldmann III (0.43 degrees diameter), color is assumed to be white. These cannot be changed.
Value
A list containing:
-
err0 all clear, >= 1 some error codes (eg cannot track, etc) (integer) -
seenFALSEfor not seen,TRUEfor seen (button pressed in response window) -
timeResponse time in ms (integer) since stimulus onset, -1 for not seen -
time_recTime since epoch when command was received at MAIA (integer ms) -
time_presTime since epoch that stimulus was presented (integer ms) -
num_track_eventsNumber of tracking events that occurred during presentation (integer) -
num_motor_failsNumber of times motor could not follow fixation movement during presentation (integer) -
pupil_diamPupil diameter in mm (float) -
loc_xX location in image of presentation (pixels, integer) -
loc_yY location in image of presentation (pixels, integer)
Examples
## Not run:
# Set up the MAIA
chooseOpi("MAIA")
result <- opiInitialize(ip = "192.168.1.7", port = 44965)
if (!is.null(result$err))
stop("Initialisation failed")
#' @param x X location of stim in degrees
#' @param y Y location of stim in degrees
#' @param size If 3, Goldmann III, else V
#' @param db Value in dB
#' @return stim object ready for opiPresent
makeStim <- function(x, y, size, db) {
s <- list(x = x, y = y, level = dbTocd(db, 10000 / pi),
size = ifelse(size == 3, 0.43, 1.77),
duration = 200, responseWindow = 1500)
class(s) <- "opiStaticStimulus"
return(s)
}
result <- opiPresent(makeStim(9, 9, 3, 10))
## End(Not run)
Implementation of opiPresent for the O600 machine.
Description
This is for internal use only. Use opiPresent() with the same arguments.
Arguments
stim |
Stimulus to present which is a list with the following elements:
|
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
Value
A list containing
-
errString message or NULL for no error. -
seen1 if seen, 0 otherwise. -
timeReaction time (if seen).
Examples
## Not run:
chooseOpi("O600")
if (!is.null(opiInitialize()$err))
stop("opiInitialize failed")
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
print(opiPresent(s, NULL))
## End(Not run)
Implementation of opiPresent for the Octopus090 machine.
Description
This is for internal use only. Use opiPresent() with the same arguments and
the class of stim as one of opiStaticStimulus, opiTemporalStimulus, or opiKineticStimulus.
Arguments
stim |
Stimulus to present (a list, see |
nextStim |
The stimulus to present after stim (it is not presented, but projector can move to it during response window) |
F310 |
If |
See Also
octo900.presentStatic, octo900.presentTemporal, octo900.presentKinetic
Examples
## Not run:
chooseOpi("Octopus900")
if (!is.null(opiInitialize()$err))
stop("opiInitialize failed")
s <- list(x=9, y=9, level=dbTocd(db), size=0.43, color="white",
duration=200, responseWindow=1500, checkFixationOK=NULL)
class(s) <- "opiStaticStimulus"
print(opiPresent(s, NULL))
## End(Not run)
Implementation of opiPresent for the PhoneHMD machine.
Description
This is for internal use only. Use opiPresent() after
chooseOPI("PhoneHMD") to call this function.
Arguments
stim |
A list containing:
|
... |
Parameters for other opiPresent implementations that are ignored here. |
Details
Elements in lum can take on values in the range [0.0, 1.0E10].
stim.length can take on values in the range [1, 2147483647].
Elements in color1 can take on values in the range [0.0, 1.0].
Elements in sx can take on values in the range [0.0, 180.0].
Elements in sy can take on values in the range [0.0, 180.0].
Elements in eye can take on values in the set
{"left", "right", "both", "none"}.
Elements in t can take on values in the range [0.0, 1.0E10].
w can take on values in the range [0.0, 1.0E10].
Elements in x can take on values in the range [-90.0, 90.0].
Elements in y can take on values in the range [-90.0, 90.0].
Elements in envSdx can take on values in the range
[-1.0E10, 1.0E10].
Elements in envSdy can take on values in the range
[-1.0E10, 1.0E10].
Elements in envRotation can take on values in
the range [-1.0E10, 1.0E10].
Elements in type can take on values in the set
{"flat", "checkerboard", "sine", "squaresine", "g1",
"g2", "g3", "text", "image"}.
Elements in frequency can take on values in the
range [0.0, 300.0].
Elements in color2 can take on values in the range [0.0, 1.0].
Elements in fullFoV can take on values in the
range [-1.0E10, 1.0E10].
Elements in phase can take on values in the range [0.0, 1.0E10].
Elements in shape can take on values in the set
{"triangle", "square", "polygon", "hollow_triangle",
"hollow_square", "hollow_polygon", "cross", "maltese",
"circle", "annulus", "optotype", "text", "model"}.
Elements in rotation can take on values in the range [0.0, 360.0].
Elements in texRotation can take on values in
the range [0.0, 360.0].
Elements in defocus can take on values in the range [0.0, 1.0E10].
Elements in envType can take on values in the set
{"none", "square", "circle", "gaussian"}.
Elements in contrast can take on values in the range [0.0, 1.0].
Elements in optotype can take on values in the
set {"a", "b", "c", "d", "e", "f", "g", "h", "i",
"j", "k", "l", "m", "n", "o", "p", "q", "r", "s", "t",
"u", "v", "w", "x", "y", "z"}.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error. -
timeResponse time from stimulus onset if button pressed (ms). -
seen'1' if seen, '0' if not.
See Also
Examples
## Not run:
chooseOpi("PhoneHMD")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiPresent(stim = list(lum = list(300.0), stim.length = 1, color1 = list(list(1.0,
1.0, 1.0)), sx = list(1.72), sy = list(1.72),
eye = list("LEFT"), t = list(200.0), w = 1500.0, x = list(0.0), y = list(0.0)))
## End(Not run)
Implementation of opiPresent for the PicoVR machine.
Description
This is for internal use only. Use opiPresent() after
chooseOPI("PicoVR") to call this function.
Arguments
stim |
A list containing:
|
... |
Parameters for other opiPresent implementations that are ignored here. |
Details
Elements in lum can take on values in the range [0.0, 1.0E10].
stim.length can take on values in the range [1, 2147483647].
Elements in color1 can take on values in the range [0.0, 1.0].
Elements in sx can take on values in the range [0.0, 180.0].
Elements in sy can take on values in the range [0.0, 180.0].
Elements in eye can take on values in the set
{"left", "right", "both", "none"}.
Elements in t can take on values in the range [0.0, 1.0E10].
w can take on values in the range [0.0, 1.0E10].
Elements in x can take on values in the range [-90.0, 90.0].
Elements in y can take on values in the range [-90.0, 90.0].
Elements in envSdx can take on values in the range
[-1.0E10, 1.0E10].
Elements in envSdy can take on values in the range
[-1.0E10, 1.0E10].
Elements in envRotation can take on values in
the range [-1.0E10, 1.0E10].
Elements in type can take on values in the set
{"flat", "checkerboard", "sine", "squaresine", "g1",
"g2", "g3", "text", "image"}.
Elements in frequency can take on values in the
range [0.0, 300.0].
Elements in color2 can take on values in the range [0.0, 1.0].
Elements in fullFoV can take on values in the
range [-1.0E10, 1.0E10].
Elements in phase can take on values in the range [0.0, 1.0E10].
Elements in shape can take on values in the set
{"triangle", "square", "polygon", "hollow_triangle",
"hollow_square", "hollow_polygon", "cross", "maltese",
"circle", "annulus", "optotype", "text", "model"}.
Elements in rotation can take on values in the range [0.0, 360.0].
Elements in texRotation can take on values in
the range [0.0, 360.0].
Elements in defocus can take on values in the range [0.0, 1.0E10].
Elements in envType can take on values in the set
{"none", "square", "circle", "gaussian"}.
Elements in contrast can take on values in the range [0.0, 1.0].
Elements in optotype can take on values in the
set {"a", "b", "c", "d", "e", "f", "g", "h", "i",
"j", "k", "l", "m", "n", "o", "p", "q", "r", "s", "t",
"u", "v", "w", "x", "y", "z"}.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error. -
timeResponse time from stimulus onset if button pressed (ms). -
seen'1' if seen, '0' if not.
See Also
Examples
## Not run:
chooseOpi("PicoVR")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiPresent(stim = list(lum = list(300.0), stim.length = 1, color1 = list(list(1.0,
1.0, 1.0)), sx = list(1.72), sy = list(1.72),
eye = list("LEFT"), t = list(200.0), w = 1500.0, x = list(0.0), y = list(0.0)))
## End(Not run)
opiPresent_for_SimGaussian
Description
Determine the response to a stimuli by sampling from a cumulative Gaussian Frequency-of-Seeing (FoS) curve (also known as the psychometric function).
The FoS has formula
\mbox{fpr}+(1-\mbox{fpr}-\mbox{fnr})(1-\mbox{pnorm}(x, \mbox{tt}, \mbox{sd})
where x is the stimulus value in dB, and sd is
set by opiInitialize and tt, fpr and fnr
are parameters.
This is for internal use only, use opiPresent().
Arguments
stim |
A list that contains at least:
|
fpr |
false positive rate for the FoS curve (range 0..1). |
fnr |
false negative rate for the FoS curve (range 0..1). |
tt |
mean of the assumed FoS curve in dB. |
... |
Any other parameters you like, they are ignored. |
Value
A list with elements:
-
errNULLif no error, a string message otherwise. -
seenTRUEorFALSE. -
timeAlwaysNA.
Examples
# Stimulus is Size III white-on-white as in the HFA
chooseOpi("SimGaussian")
res <- opiInitialize(sd = 1.6)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$err))
result <- opiPresent(stim = list(level = dbTocd(20)), tt = 30, fpr = 0.15, fnr = 0.01)
print(paste("Seen:", result$seen, quote = FALSE))
res <- opiClose()
if (!is.null(res$err))
warning(paste("opiClose() failed:", res$err))
opiPresent_for_SimHenson
Description
Determine the response to a stimuli by sampling from a cumulative Gaussian Frequency-of-Seeing (FoS) curve (also known as the psychometric function).
For internal use only, use opiPresent().
Arguments
stim |
A list that contains at least:
|
fpr |
false positive rate for the FoS curve (range 0..1). |
fnr |
false negative rate for the FoS curve (range 0..1). |
tt |
mean of the assumed FoS curve in dB. |
... |
Any other parameters you like, they are ignored. |
Details
The FoS formula is
\mbox{fpr}+(1-\mbox{fpr}-\mbox{fnr})(1-\mbox{pnorm}(x, \mbox{tt}, \mbox{pxVar})
where x is the stimulus value in dB, and pxVar is
\min(\mbox{cap}, e^{A\times\mbox{tt}+B}).
The ceiling cap is set with the call to
opiInitialize, and A and B are from Table 1 in Henson
et al (2000), also set in the call to opiInitialise using the type parameter.
Value
A list with elements:
-
errNULLif no error, a string message otherwise. -
seenTRUEorFALSE. -
timeAlwaysNA.
Examples
# Stimulus is Size III white-on-white as in the HFA
chooseOpi("SimHenson")
res <- opiInitialize(type = "C", cap = 6)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$err))
result <- opiPresent(stim = list(level = dbTocd(20)), tt = 30, fpr = 0.15, fnr = 0.01)
print(paste("Seen:", result$seen, quote = FALSE))
res <- opiClose()
if (!is.null(res$err))
stop(paste("opiClose() failed:", res$err))
opiPresent_for_SimHensonRT
Description
Determine the response to a stimuli by sampling from a cumulative Gaussian Frequency-of-Seeing (FoS) curve (also known as the psychometric function).
For internal use only, use opiPresent().
Arguments
stim |
A list that contains at least:
|
fpr |
false positive rate for the FoS curve (range 0..1). |
fnr |
false negative rate for the FoS curve (range 0..1). |
tt |
mean of the assumed FoS curve in dB. |
dist |
The distance of the stimulus level from |
... |
Any other parameters you like, they are ignored. |
Details
As the response time returned for a false positive is determined separately from a positive response, we first check for a false response. If there is no false response, we use the FoS formula
1-\mbox{pnorm}(x, \mbox{tt}, \mbox{pxVar})
where x is the stimulus value in dB, and pxVar is
\min(\mbox{cap}, e^{A\times\mbox{tt}+B}).
The ceiling cap is set with the call to
opiInitialize, and A and B are from Table 1 in Henson
et al (2000), also set in the call to opiInitialise using the type parameter.
Thus, this function is the same as for SimHenson, but
reaction times are determined by sampling from rtData as passed to
opiInitialize. The dist parameter is the distance of the
stimulus level from the true threshold, and should be in the same units as
the Dist column of rtData. The default is just the straight
difference between the stimulus level and the true threshold, but you might
want it scaled somehow to match rtData.
Value
A list with elements:
-
errNULLif no error, a string message otherwise. -
seenTRUEorFALSE. -
timeThe response time.
Examples
# Stimulus is Size III white-on-white as in the HFA
chooseOpi("SimHensonRT")
data(RtSigmaUnits)
res <- opiInitialize(type = "C", cap = 6, rtData = RtSigmaUnits)
if (!is.null(res$err))
stop(paste("opiInitialize() failed:", res$err))
dist <- (10 - 30) / min(exp(-0.098 * 30 + 3.62), 6)
result <- opiPresent(stim = list(level = dbTocd(20)), tt = 30, fpr = 0.15, fnr = 0.01, dist=dist)
print(result, quote = FALSE)
if (!is.null(opiClose()))
warning("opiClose() failed")
opiPresent_for_SimNo
Description
Always respond 'not seen' to any parameter.
No checking is done on the validity of stim.
Arguments
stim |
Anything you like, it is ignored. |
... |
Any parameters you like, they are ignored. |
Value
A list with elements:
-
errAlwaysNULL. -
timeAlwaysNA.
opiPresent_for_SimYes
Description
Always respond 'yes' immediately to any parameter.
Arguments
stim |
Anything you like, it is ignored. |
... |
Any parameters you like, they are ignored. |
Value
A list with elements:
-
errAlwaysFALSE. -
seenAlwaysTRUE. -
timeAlwaysNA.
Calls opiQueryDevice_for_MACHINE as appropriate.
Description
Returns a list that describes the current state of the machine. Specific parameters and return values can be seen in the machine specific versions listed below in the ’See Also’.
Usage
opiQueryDevice()
Value
A list specific to each machine.
See Also
opiQueryDevice_for_ImoVifa(),
opiQueryDevice_for_Compass(),
opiQueryDevice_for_Octopus900(),
opiQueryDevice_for_PhoneHMD(), opiQueryDevice_for_Display(), opiQueryDevice_for_PicoVR(),
opiQueryDevice_for_SimNo(), opiQueryDevice_for_SimYes(), opiQueryDevice_for_SimHenson(),
opiQueryDevice_for_SimHensonRT(),
opiQueryDevice_for_SimGaussian()
Implementation of opiQueryDevice for the ImoVifa machine.
Description
This is for internal use only. Use opiQueryDevice() with
these Arguments and you will get the Value back.
Value
A list containing constants and their values used in the OPI Compass module.
Implementation of opiQueryDevice for the Display machine.
Description
This is for internal use only. Use opiQueryDevice() after
chooseOPI("Display") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("Display")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiQueryDevice()
## End(Not run)
Implementation of opiQueryDevice for the ImoVifa machine.
Description
This is for internal use only. Use opiQueryDevice() after
chooseOPI("ImoVifa") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error. -
rightEyexx co-ordinates of right pupil (pixels from image centre, left < 0) -
rightEyeyy co-ordinates of right pupil (pixels from image centre, up > 0) -
leftEyexx co-ordinates of left pupil (pixels from image centre, left < 0) -
leftEyeyy co-ordinates of left pupil (pixels from image centre, up > 0) -
leftEyedDiameter of left pupil (pixels) -
rightEyedDiameter of right pupil (pixels)
See Also
Examples
## Not run:
chooseOpi("ImoVifa")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiQueryDevice()
## End(Not run)
Implementation of opiQueryDevice for the Kowa AP7000 machine.
Description
This is for internal use only. Use opiQueryDevice() with
these Arguments and you will get the Value back.
Details
Prints defined constants in OPI package pertaining to Kowa AP7000.
Value
List containing
-
isSim = FALSE. -
pupilX, the x-coordinate of the pupil position in pixels. -
pupilY, the y-coordinate of the pupil position in pixels. -
purkinjeX, the x-coordinate of the purkinje position in pixels. -
purkinjeY, the y-coordinate of the purkinje position in pixels.
Implementation of opiQueryDevice for the ImoVifa machine.
Description
This is for internal use only. Use opiQueryDevice() with
these Arguments and you will get the Value back.
Value
A list containing constants and their values used in the OPI MAIA module.
Implementation of opiQueryDevice for the O600 machine.
Description
This is for internal use only. Use opiQueryDevice() with
these Arguments and you will get the Value back.
Details
Prints defined constants in OPI package pertaining to O600.
Value
Returns a list of 10 items:
* \code{answerButton} 0 = not pressed, 1 = pressed
* \code{headSensor} 0 = no forehead detected, 1 = forehead detected
* \code{eyeLidClosureLeft} 0 = eye is open, 1 = eye is closed
* \code{eyeLidClosureRight} 0 = eye is open, 1 = eye is closed
* \code{fixationLostLeft} 1 = eye pos lost, 0 = eye pos ok
* \code{fixationLostRight} 1 = eye pos lost, 0 = eye pos ok
* \code{pupilPositionXLeft} (in px)
* \code{pupilPositionYLeft} (in px)
* \code{pupilPositionXRight} (in px)
* \code{pupilPositionYRight} (in px)
Implementation of opiQueryDevice for the Octopus900 machine.
Description
This is for internal use only. Use opiQueryDevice() with
these Arguments and you will get the Value back.
Details
Prints defined constants in OPI package pertaining to Octopus 900.
Value
List containing isSim = FALSE.
Implementation of opiQueryDevice for the PhoneHMD machine.
Description
This is for internal use only. Use opiQueryDevice() after
chooseOPI("PhoneHMD") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PhoneHMD")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiQueryDevice()
## End(Not run)
Implementation of opiQueryDevice for the PicoVR machine.
Description
This is for internal use only. Use opiQueryDevice() after
chooseOPI("PicoVR") to call this function.
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PicoVR")
opiInitialise(list(port = 50001, ip = "localhost"))
opiSetup(list(eye = "BOTH"))
result <- opiQueryDevice()
## End(Not run)
opiQueryDevice_for_SimGaussian
Description
Returns a simple list.
Value
A list with elements:
-
errAlwaysNULL -
machinethat is set to"SimGaussian".
opiQueryDevice_for_SimHenson
Description
Returns name of the machine.
Value
A list with elements:
-
isSimAlwaysTRUE. -
machinethat is set to"SimHenson".
opiQueryDevice_for_SimHensonRT
Description
Returns name of the machine.
Value
A list with elements:
-
isSimAlwaysTRUE. -
machinethat is set to"SimHensonRT".
opiQueryDevice_for_SimNo
Description
Returns name of the machine.
Value
A list with elements:
-
machinethat is set to"SimNo". -
isSimthat is set to TRUE.
opiQueryDevice_for_SimYes
Description
Returns name of the machine.
Value
A list with elements:
-
isSimAlwaysTRUE. -
machinethat is set to"SimYes".
Deprecated. Use opiSetup().
Description
In older OPIs it set background color and luminance in both eyes.
Deprecated for OPI >= v3.0.0 and replaced with opiSetup().
See Also
Calls opiSetup_for_MACHINE as appropriate.
Description
Specific parameters and return values can be seen in the machine specific versions listed below in the ’See Also’.
Usage
opiSetup(settings)
Arguments
settings |
A list containing specific settings for a machine. |
Value
Each implementation should(!) return a list with at least the following elements:
-
errNULLif no error, otherwise a string describing the error.
See Also
opiSetup_for_Compass(),
opiSetup_for_Octopus900(),
opiSetup_for_ImoVifa(),
opiSetup_for_PhoneHMD(), opiSetup_for_Display(), opiSetup_for_PicoVR(),
opiSetup_for_SimNo(), opiSetup_for_SimYes(), opiSetup_for_SimHenson(),
opiSetup_for_SimHensonRT(),
opiSetup_for_SimGaussian()
Implementation of opiSetup for the Compass machine.
Description
This is for internal use only. Use opiSetup() with
these Arguments and you will get the Value back.
Arguments
settings |
is a list that could contain:
|
Details
Note: tracking will be relative to the PRL established with the fixation marker used at setup (call to OPI-OPEN), so when tracking is on you should use the same fixation location as in the setup.
Value
A list containing err which is NULL for success, or some string description for fail.
Implementation of opiSetup for the Display machine.
Description
This is for internal use only. Use opiSetup() after
chooseOPI("Display") to call this function.
Arguments
settings |
A list containing:
|
Details
eye can take on values in the set {"left", "right", "both",
"none"}.
fixShape can take on values in the set {"triangle",
"square", "polygon", "hollow_triangle", "hollow_square",
"hollow_polygon", "cross", "maltese", "circle", "annulus",
"optotype", "text", "model"}.
fixLum can take on values in the range [0.0, 1.0E10].
fixType can take on values in the set {"flat",
"checkerboard", "sine", "squaresine", "g1", "g2", "g3", "text", "image"}.
fixCx can take on values in the range [-90.0, 90.0].
fixCy can take on values in the range [-90.0, 90.0].
Elements in fixCol can take on values in the range [0.0, 1.0].
bgLum can take on values in the range [0.0, 1.0E10].
tracking can take on values in the range [0, 1].
Elements in bgCol can take on values in the range [0.0, 1.0].
fixSx can take on values in the range [0.0, 1.0E10].
fixSy can take on values in the range [0.0, 1.0E10].
fixRotation can take on values in the range [0.0, 360.0].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("Display")
opiInitialise(list(port = 50001, ip = "localhost"))
result <- opiSetup(settings = list(eye = "BOTH"))
## End(Not run)
Implementation of opiSetup for the ImoVifa machine.
Description
This is for internal use only. Use opiSetup() after
chooseOPI("ImoVifa") to call this function.
Arguments
settings |
A list containing:
|
Details
eye can take on values in the set {"left", "right", "both",
"none"}.
fixShape can take on values in the set {"triangle",
"square", "polygon", "hollow_triangle", "hollow_square",
"hollow_polygon", "cross", "maltese", "circle", "annulus",
"optotype", "text", "model"}.
fixLum can take on values in the range [0.0, 1.0E10].
fixType can take on values in the set {"flat",
"checkerboard", "sine", "squaresine", "g1", "g2", "g3", "text", "image"}.
fixCx can take on values in the range [-90.0, 90.0].
fixCy can take on values in the range [-90.0, 90.0].
Elements in fixCol can take on values in the range [0.0, 1.0].
bgLum can take on values in the range [0.0, 1.0E10].
tracking can take on values in the range [0, 1].
Elements in bgCol can take on values in the range [0.0, 1.0].
fixSx can take on values in the range [0.0, 1.0E10].
fixSy can take on values in the range [0.0, 1.0E10].
fixRotation can take on values in the range [0.0, 360.0].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("ImoVifa")
opiInitialise(list(port = 50001, ip = "localhost"))
result <- opiSetup(settings = list(eye = "BOTH"))
## End(Not run)
opiSetBackground
Description
Implementation of opiSetup for the Kowa AP7000 machine.
This is for internal use only. Use opiSetup() with these Arguments.
Usage
opiSetup_for_KowaAP7000(lum = NA, color = NA, fixation = NA)
Arguments
lum |
Must be 10 for a white background and 100 for a yellow. |
color |
One of |
fixation |
One of
* |
Details
If lum is 10 and color is not set, then
.opi_env$KowaAP7000$BACKGROUND_WHITE is assumed.
If lum is 100 and color is not set,
then .opi_env$KowaAP7000$BACKGROUND_YELLOW is assumed.
If both lum and color is set, then lum is ignored
(a warning will be generated
If lum is incompatible with color).
Value
Always returns list(err = NULL)
Implementation of opiSetup for the MAIA machine.
Description
This is for internal use only. Use opiSetup() with
these Arguments and you will get the Value back.
Arguments
settings |
is a list that could contain:
|
Details
Note that corner locations are missing, so both x and y cannot be +-4.8.
Value
A list containing
* err which is NULL for success
* message for some messages from MAIA (probably "0")
For open list might (but might not) contain:
-
prlA pair giving the (x,y) in degrees of the Preferred Retinal Locus detected in the initial alignment. -
onha pair giving the (x,y) in degrees of the ONH as selected by the user. -
imageraw bytes being the JPEG compressed infra-red image acquired during alignment.
opiSetup for the O600 machine.
Description
Implementation of opiSetup for the O600 machine.
This is for internal use only. Use opiSetup() with these Arguments.
Arguments
bgColor |
Background color 0 to 255. |
fixType |
fixation type 1, 2, 3 or 4. |
fixColor |
fixation color 0, 2, 4 or 5. |
fixIntensity |
fixation point intensity 0 to 255. |
Value
A list containing one of the following
-
err = -1to be implemented -
err = -2O600 sent back an error; bad background parameters -
err = -3O600 sent back an error; bad fixation parameters -
err = NULLSuccess
Implementation of opiSetup for the Octopus900 machine.
Description
This is for internal use only.
Use opiSetup() with the same parameters.
Arguments
lum |
Luminance level in cd/m^2 |
color |
Stimulus color (see details) |
fixation |
fixation target |
fixIntensity |
fixation point intensity |
Details
Allowable lum and color are defined in the .opi_env$O900 environment.
-
lumis intensity of the background and can be one of-
.opi_env$O900$BG_OFF, which turns background off. -
.opi_env$O900$BG_1, background of 1.27 cd/\mbox{m}^2. -
.opi_env$O900$BG_10, background of 10 cd/\mbox{m}^2. -
.opi_env$O900$BG_100, background of 100 cd/\mbox{m}^2.
-
-
colorcan be one of the following choices.-
.opi_env$O900$MET_COL_WWfor white-on-white -
.opi_env$O900$MET_COL_RWfor red-on-white -
.opi_env$O900$MET_COL_BWfor blue-on-white -
.opi_env$O900$MET_COL_WYfor white-on-yellow -
.opi_env$O900$MET_COL_RYfor red-on-yellow -
.opi_env$O900$MET_COL_BYfor blue-on-yellow
-
-
fixationis one of-
.opi_env$O900$FIX_CENTREor.opi_env$O900$FIX_CENTER -
.opi_env$O900$FIX_CROSS -
.opi_env$O900$FIX_RING
-
-
fixIntensityis a percentage between 0 and 100. 0 is off, 100 the brightest.
Note if you specify fixation you also have to specify fixIntensity.
Value
A list with element err which is
NULL on success
-1 indicates
opiInitializehas not been called.-2 indicates could not set the background color.
-3 indicates could not set the fixation marker.
or a string message about bad parameters
Examples
## Not run:
chooseOpi("Octopus900")
oi <- opiInitialize(eyeSuiteJarLocation="c:/EyeSuite/",
eyeSuiteSettingsLocation="c:/Documents and Settings/All Users/Haag-Streit/",
eye="left")
if(!is.null(oi$err))
stop("opiInitialize failed")
if(!is.null(opiSetup(fixation=.opi_env$O900$FIX_CENTRE)$err))
stop("opiSetup failed")
if(!is.null(opiSetup(fixation=.opi_env$O900$FIX_RING, fixIntensity=0)$err))
stop("opiSetup failed")
if(!is.null(opiSetup(color=.opi_env$O900$MET_COL_BY)$err))
stop("opiSetup failed")
if(!is.null(opiSetup(lum=.opi_env$O900$BG_100, color=.opi_env$O900$MET_COL_RW)$err))
stop("opiSetup failed")
opiClose()
## End(Not run)
Implementation of opiSetup for the PhoneHMD machine.
Description
This is for internal use only. Use opiSetup() after
chooseOPI("PhoneHMD") to call this function.
Arguments
settings |
A list containing:
|
Details
eye can take on values in the set {"left", "right", "both",
"none"}.
fixShape can take on values in the set {"triangle",
"square", "polygon", "hollow_triangle", "hollow_square",
"hollow_polygon", "cross", "maltese", "circle", "annulus",
"optotype", "text", "model"}.
fixLum can take on values in the range [0.0, 1.0E10].
fixType can take on values in the set {"flat",
"checkerboard", "sine", "squaresine", "g1", "g2", "g3", "text", "image"}.
fixCx can take on values in the range [-90.0, 90.0].
fixCy can take on values in the range [-90.0, 90.0].
Elements in fixCol can take on values in the range [0.0, 1.0].
bgLum can take on values in the range [0.0, 1.0E10].
tracking can take on values in the range [0, 1].
Elements in bgCol can take on values in the range [0.0, 1.0].
fixSx can take on values in the range [0.0, 1.0E10].
fixSy can take on values in the range [0.0, 1.0E10].
fixRotation can take on values in the range [0.0, 360.0].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PhoneHMD")
opiInitialise(list(port = 50001, ip = "localhost"))
result <- opiSetup(settings = list(eye = "BOTH"))
## End(Not run)
Implementation of opiSetup for the PicoVR machine.
Description
This is for internal use only. Use opiSetup() after
chooseOPI("PicoVR") to call this function.
Arguments
settings |
A list containing:
|
Details
eye can take on values in the set {"left", "right", "both",
"none"}.
fixShape can take on values in the set {"triangle",
"square", "polygon", "hollow_triangle", "hollow_square",
"hollow_polygon", "cross", "maltese", "circle", "annulus",
"optotype", "text", "model"}.
fixLum can take on values in the range [0.0, 1.0E10].
fixType can take on values in the set {"flat",
"checkerboard", "sine", "squaresine", "g1", "g2", "g3", "text", "image"}.
fixCx can take on values in the range [-90.0, 90.0].
fixCy can take on values in the range [-90.0, 90.0].
Elements in fixCol can take on values in the range [0.0, 1.0].
bgLum can take on values in the range [0.0, 1.0E10].
tracking can take on values in the range [0, 1].
Elements in bgCol can take on values in the range [0.0, 1.0].
fixSx can take on values in the range [0.0, 1.0E10].
fixSy can take on values in the range [0.0, 1.0E10].
fixRotation can take on values in the range [0.0, 360.0].
Value
A list containing:
-
errNULLif there was no error, a string message if there is an error.
See Also
Examples
## Not run:
chooseOpi("PicoVR")
opiInitialise(list(port = 50001, ip = "localhost"))
result <- opiSetup(settings = list(eye = "BOTH"))
## End(Not run)
opiSetup_for_SimGaussian
Description
Does nothing.
Arguments
... |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
opiSetup_for_SimHenson
Description
Does nothing.
For internal use only, use opiSetup().
Arguments
... |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
opiSetup_for_SimHensonRT
Description
Does nothing.
For internal use only, use opiSetup().
Arguments
... |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
opiSetup_for_SimNo
Description
Does nothing.
Arguments
settings |
Anything you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
opiSetup_for_SimYes
Description
Does nothing.
Arguments
settings |
Any object you like, it is ignored. |
Value
A list with elements:
-
errAlwaysNULL.
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Description
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Usage
opiStaticStimulus()
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Description
For backwards compatibility. Used by Octopus900 and KowaAP7000.
Usage
opiTemporalStimulus()
Convert pixels to degrees for machine 'machine'
Description
Convert pixels to degrees for machine 'machine'
Usage
pixTodeg(xy, machine = "compass")
Arguments
xy |
a 2 element vector c(x,y) where x and y are in pixels |
machine |
"compass" or ...? |
Value
xy converted to degrees of visual field with the usual conventions or NA if machine is unknown
Examples
pixTodeg(c(1000, 200), machine="compass") # c(1.290323, 24.516129) degrees